8.2 Shutdown
Before maintenance or storage, remove power to the instrument and stop the flow of sample water
and cooling water to the instrument. Refer to the illustrated steps in
Note: If cooling water is not connected to the instrument, there is only one shutoff valve.
Figure 10 Set the system to off
8.3 Clean the sensors
1.
Examine the sensor(s) periodically for unwanted material.
2.
Clean the sensor when unwanted material collects on the surface of the electrode or when
performance decreases. Refer to the Polymetron 8315 sensor documentation.
3.
If a sensor shows low performance after cleaning, replace the sensor.
English
23