The 9525sc DCCP system removes CO
2
and measures specific conductivity (SC), cation
conductivity (CC) and degassed cation conductivity (DCC) and calculates pH. The system measures
a sample stream of high purity water. The system has three conductivity sensors and two controllers.
The top controller is connected to two sensors and shows the SC (channel 1) and CC (channel 2)
measurements. The bottom controller shows the DCC measurement. Refer to
.
Figure 2 9525sc DCCP system
1
Heater
7
Power switch
2
Sample cooler
8
Power indicator light
2
3
SC200 Controller (or Polymetron
9500 controller)
9
Temperature controller
4
Polymetron 8315 conductivity sensor (3x)
10
Sample flow meter
5
Electrical enclosure
11
Sample flow valve
6
Flow cell (3x)
12
Resin column
2
The power indicator light starts and stops with the heater.
8
English