48
Maintenance
6.3 Routine maintenance schedule
The maintenance schedule is given for standard applications.
Deviant applications may cause different maintenance intervals.
Table 6 Routine maintenance schedule
Description
3 months
(Customer
task)
6 months
(Service
task)
12 months
(Service
task)
24 months
(Service
task)
Visual check analytical compartment, manual cleaning if
necessary.
X
1
X
Check filter pads, clean/ replace if necessary, particularly on
fan side.
X
1
X
Check reagents, replace if necessary.
X
1
X
Check cleaning solution, replace if necessary.
X
1
X
Check maintenance counters.
X
1
X
Check standard solutions, replace if necessary.
X
1
X
Visual check of electrode, replace membrane cap and
electrolyte if necessary
X
1
X
Check system for air impermeability.
X
Function check of both fans.
X
Function check heating for analyser enclosure.
X
General function check.
X
Read out and analyse event log. Read out and check data log
if necessary.
X
Check electrode (slope with intact membrane cap: –55mV to
–67mV), check every 6 months after 12 months of use.
(X)
2
X
Replace pump head for air pump.
X
Check magnetic stir bar, replace if necessary.
X
Replace reagent pump.
X
Check cleaning pump and replace if necessary (check every
6 months after 12 months of use)
(X)
2
X
Check stirring motor, replace if necessary.
X
1
Recommended maintenance interval, especially for reagents. The actual reagent and electrolyte (AMTAX sc) exchange
intervals depend on configuration.
2
Maintenance cycles are given for standard applications. Deviant applications may cause different maintenance intervals.
Summary of Contents for AMTAX sc
Page 2: ......
Page 7: ...7 Specifications Figure 1 Instrument dimensions AMTAX sc...
Page 12: ...12 General Information...
Page 36: ...36 System Start Up...
Page 44: ...44 Operation...
Page 56: ...56 Maintenance...
Page 64: ...64 Troubleshooting...
Page 69: ...69 Replacement Parts and Accessorie Figure 23 Analyzer enclosure AMTAX indoor sc...
Page 70: ...70 Replacement Parts and Accessorie Figure 24 Analyzer panel overview...
Page 71: ...71 Replacement Parts and Accessorie Figure 25 Analyzer panel front view details...
Page 72: ...72 Replacement Parts and Accessorie Figure 26 Analyzer panel back view details...
Page 76: ...76 Warranty and liability...