92
Plumbing and Connection Options
A.11 Option 7 plumbing and connections
Option 7 is used with an sc analyzer and the Filter Probe sc. The
waste from the analyzer is discharged back into the basin using the
Filtration Kit. Use the drain tube inside the Filter Probe sc or the
optional heated drain tube to discharge the waste stream from the
sc analyzer.
Refer to
and the following instructions for Option 7:
1.
Install the Filter Probe sc into the sample stream. Refer to the
Filter Probe sc User Manual for more information.
2.
Feed the Filter Probe sc hose (sample lines, electrical cables,
drain tube) through the analyzer opening (
Use Seal Plug #2 to secure.
3.
Seal the unused openings with Seal Plug #3.
4.
Connect the Filter Probe sc data cable and power connections.
Refer to
.
5.
Connect the air tube to the compressor (item 4).
6.
Connect the drain tube and feed it out the analyzer through
Seal Plug #3 to a drain.
7.
Connect the sample line to the sample inlet on the overflow
vessel using the fittings (item 8).
Summary of Contents for AMTAX sc
Page 2: ......
Page 7: ...7 Specifications Figure 1 Instrument dimensions AMTAX sc...
Page 12: ...12 General Information...
Page 36: ...36 System Start Up...
Page 44: ...44 Operation...
Page 56: ...56 Maintenance...
Page 64: ...64 Troubleshooting...
Page 69: ...69 Replacement Parts and Accessorie Figure 23 Analyzer enclosure AMTAX indoor sc...
Page 70: ...70 Replacement Parts and Accessorie Figure 24 Analyzer panel overview...
Page 71: ...71 Replacement Parts and Accessorie Figure 25 Analyzer panel front view details...
Page 72: ...72 Replacement Parts and Accessorie Figure 26 Analyzer panel back view details...
Page 76: ...76 Warranty and liability...