26
Installation
3.5.3 Insert tubing and/or cables
3.5.4 Connect the Filter Probe to the analyzer
CAUTION
The enclosure may tip forwards if it has not been fixed in
place. Only open the enclosure if the enclosure is properly
mounted.
DANGER
Disconnect power from the sc analyzer at the sc1000 before
removing the protective covers in the analyzer.
Important Note:
Make sure that for the use of a filter probe, the
level of the water in which the filtration probe is submersed, is
below the level of the analyzer.
1.
Open the enclosure door and secure with the door hook.
2.
Open the analyzer panel.
3.
Remove the two screws from the protective cover and remove
the cover (item 1,
).
4.
Connect the earth ground (green/yellow) wire (item 9,
) from the Filter Probe sc to the ground
terminal (item 5,
5.
Connect the power connector to the appropriate terminal
connection (items 4 and 11,
).
6.
Remove the three screws securing the bottom panel cover.
(item 2,
). Remove the panel.
7.
Connect the data connector (item 10,
) to
the main board (item 12,
8.
Install all covers and panels.
9.
Connect the white air tubing (item 8,
) from
the Filter Probe sc to the air tubing connection on the analyzer
(
).
10.
For sample and drain line connections refer to
1
Guide the tubing or cables through
the enclosure openings
(
).
2
Push the plug from the top onto the
tubing or onto the cable.
3
Pull down the plug with the tubing
or the cables. Seal any unused
entry with Plug #3.
Summary of Contents for AMTAX sc
Page 2: ......
Page 7: ...7 Specifications Figure 1 Instrument dimensions AMTAX sc...
Page 12: ...12 General Information...
Page 36: ...36 System Start Up...
Page 44: ...44 Operation...
Page 56: ...56 Maintenance...
Page 64: ...64 Troubleshooting...
Page 69: ...69 Replacement Parts and Accessorie Figure 23 Analyzer enclosure AMTAX indoor sc...
Page 70: ...70 Replacement Parts and Accessorie Figure 24 Analyzer panel overview...
Page 71: ...71 Replacement Parts and Accessorie Figure 25 Analyzer panel front view details...
Page 72: ...72 Replacement Parts and Accessorie Figure 26 Analyzer panel back view details...
Page 76: ...76 Warranty and liability...