27
Installation
3.5.5 Connect the optional heated drain
CAUTION
The enclosure may tip forwards if it has not been fixed in
place. Only open the enclosure if the enclosure is properly
mounted.
DANGER
Disconnect power from the sc analyzer at the sc1000 before
removing the protective covers in the analyzer.
Refer to
and the following procedure to connect the
heated drain.
1.
Open the enclosure door and secure if necessary.
2.
Open the analyzer panel.
3.
Remove the protective cover (
).
4.
Connect the earth ground wire (green/yellow) to the ground
wire terminal strip.
5.
Connect the cables for the heated drain (item 6,
) to the terminal block (item 3,
).
6.
Connect the drain tube as described in the appropriate Option
Configuration, refer to
for more
information.
7.
Place the drain tube to the appropriate drain or basin.
8.
Install all covers and panels.