37
Section 5
Operation
The AMTAX sc can only be operated with an sc1000 controller. For
further information, see the sc1000 User Manual.
An LED on the door indicates the current operating state. Refer to
the sc1000 User Manual and
Instrument, chemicals and electrode are temperature-sensitive. To
prevent incorrect measurements, only operate the instrument with
the door closed.
5.1 Sensor diagnostics menu
SELECT AMTAX sc (if more than one sensor or analyzer is attached)
AMTAX sc
ERROR LIST
Displays all errors currently present in the sensor
WARNING LIST
Displays all warnings currently present in the sensor
5.2 Sensor setup menu
SELECT AMTAX sc (if more than one sensor or analyzer is attached)
CALIBRATION (see
CORR. FACTOR
Displays the correction factor
LOCATION 1
Displays location1 which is adjusted in the CONFIGURE menu
GAIN CORR
Adjusts the correction factor for channel 1
LOCATION2
Available with the 2-channel version.
GAIN CORR
Available with the 2-channel version.
CALIBRATE
Triggers an automatic calibration, then measuring mode
CAL.-CLEAN
Triggers an automatic calibration with subsequent automatic cleaning, then measuring mode.
SETTINGS
SET OUTMODE
Value that is output during a calibration and the following discarded values.
HOLD = last measured value, SET TRANSFER = value to be entered.
AUTOCAL
SET INTERVAL
Time interval between two calibrations
START
Start time for the calibration (In case of more than one calibration per day: Enter start time for
the first calibration.)
DISCHARGE
Number of measured values that are discarded after a calibration.
SEL ADJ METHOD
Select the calibration method
ADVANCED
Default calibration method for software version
≥
1.60 (For best accuracy with lowest
measuring values.)
CONVENTIONAL
Calibration method for software version
<
1.60 is available for compatiblility reasons (Not
selectable in the measuring range 0.02–5 mg/L.).
DEFAULT SETUP
Resets to factory default settings.
CONFIGURE
LOCATION 1
Settings for location 1
EDIT NAME
Enter the name for the measuring location as required.
SET PARAMETER
Select output: ammonium or ammonium nitrogen
SELECT UNITS
Select output: mg/L or ppm
QUANTITY CH 1
Number of measurements in succession ( = measurements on channel 1 +
DISCHARGE VAL 1 channel 1). Available with the 2-channel version.
Summary of Contents for AMTAX sc
Page 2: ......
Page 7: ...7 Specifications Figure 1 Instrument dimensions AMTAX sc...
Page 12: ...12 General Information...
Page 36: ...36 System Start Up...
Page 44: ...44 Operation...
Page 56: ...56 Maintenance...
Page 64: ...64 Troubleshooting...
Page 69: ...69 Replacement Parts and Accessorie Figure 23 Analyzer enclosure AMTAX indoor sc...
Page 70: ...70 Replacement Parts and Accessorie Figure 24 Analyzer panel overview...
Page 71: ...71 Replacement Parts and Accessorie Figure 25 Analyzer panel front view details...
Page 72: ...72 Replacement Parts and Accessorie Figure 26 Analyzer panel back view details...
Page 76: ...76 Warranty and liability...