Plumbing and Connection Options
A.2.1 Remove the T-fitting
When using the 2-parameter configuration, the T-fitting from the
first analyzer drain tube must be removed and reused to connect
the drain tube from the first analyzer to the second analyzer.
The T-fitting is used to connect the drain tube. To remove the
T-fitting refer to
and the following steps:
1.
Remove the drain tube from both ends of the T-fitting.
2.
Remove the drain tube assembly.
3.
Reconnect the T-fitting as described in Options 4, 6, 8b, 9b,
10b and 11b.
A.3 Drain line considerations
Always place the drain tubing so that there is a continuous fall (min.
3 degrees) and the outlet is clear (not pressurized). Make sure the
drain tubing is no longer than 2 meters.
A.4 Tubing considerations
The AMTAX sc uses four different tubing types for plumbing
connections. The type of tubing used depends on the system
configuration option:
•
Ø 3.2 mm: sample line tubing
•
Ø 6 mm: unheated drain tubing
•
Ø 22 mm: heated drain tubing
•
Ø 32 mm: tubing for the Filter Probe sc
Figure 28 Removing the T-fitting
1
T-fitting to be removed
Summary of Contents for AMTAX sc
Page 2: ......
Page 7: ...7 Specifications Figure 1 Instrument dimensions AMTAX sc...
Page 12: ...12 General Information...
Page 36: ...36 System Start Up...
Page 44: ...44 Operation...
Page 56: ...56 Maintenance...
Page 64: ...64 Troubleshooting...
Page 69: ...69 Replacement Parts and Accessorie Figure 23 Analyzer enclosure AMTAX indoor sc...
Page 70: ...70 Replacement Parts and Accessorie Figure 24 Analyzer panel overview...
Page 71: ...71 Replacement Parts and Accessorie Figure 25 Analyzer panel front view details...
Page 72: ...72 Replacement Parts and Accessorie Figure 26 Analyzer panel back view details...
Page 76: ...76 Warranty and liability...