PAT700 Total Organic Carbon Analyzer – Specifications
Page 141
Anatel
Operator Manual
Appendix A: Specifications
PAT700 specifications are subject to change without notice.
TOC specifications
Operating range
0.5 to 2,000 ppb as carbon
Display resolution
0.1 ppb
Accuracy
±1 ppb or 5%, whichever is greater
Repeatability
±0.3 ppb or ±1%, whichever is greater
Limit of detection
0.5 ppb
Conductivity and resistivity specifications
Conductivity range (TOC)
0.05 to 150 µS/cm (@ 25 °C)
Conductivity display resolution
0.01 µS/cm
Conductivity accuracy
±1% over full range (uncompensated)
Resistivity range
0.2 to 18 M
Ω
-cm (@25 °C)
Resistivity display resolution
0.01 over full range
Available resistivity modes
Temperature compensated to 25 °C, or uncompensated
Temperature specifications
Ambient operating range
10 to 40° C (50 to 104 °F)
Display resolution
0.1 over full range
Measurement accuracy
±0.4°C
Sample water range
5 to 95 °C (41 to 203 °F)