PAT700 Total Organic Carbon Analyzer – Bottle Mode
Page 97
Anatel
Operator Manual
Figure 6-24 Excursion mode dialog box
•
Excursion with validation button
opens the excursion with validation load bottles
dialog box. See
and follow the instructions in
below.
•
Excursion without validation
button tells the analyzer to run the excursion without
validation and opens the excursion without validation load bottles dialog box. See
and follow the instructions in
Excursion without validation, page 100
•
Return icon
returns you to the bottle mode dialog box.
Excursion with validation
CAUTION
Potential puncture hazard! Contact with the exposed needle could cause serious
injury. Use extreme care when installing and removing the sample bottles.
1)
Install the standards bottles following the instructions as shown in the load bottles
dialog box and ensuring that you place the standards bottle in the correct position.
See
. From left:
•
Bottle 1 = blank
•
Bottle 2 = empty
•
Bottle 3 = 500 ppb sucrose
•
Bottle 4 = excursion bottle