Setup – PAT700 Total Organic Carbon Analyzer
Operator Manual
Anatel
WG
M - 2
8
No
vemb
e
r 20
07
- Ed
it
io
n 0
1
•
A high to low state transition on digital1 aborts the current operation and immediately
runs a single TOC analysis. The analysis runs to completion regardless of whether
subsequent trigger values are received. Once the analysis is complete, the analyzer
returns to the mode defined by digital2.
•
If digital2 is high, the analyzer enters online TOC analysis mode. If digital2 is low, the
analyzer enters conductivity mode.
Average last results checkbox
averages the last X TOC analyses. The averaged value is
displayed on the home screen and in the log view, logged to the data log, and sent to the
printer based on the printer settings. The default value is disabled.
Return icon
returns you to the setup dialog box and saves the changes. Any changes are
noted in the audit trail. All changes take effect on the next change of state.
Idle tab
Figure 5-3 TOC dialog box, idle tab
Flow during idle radio buttons
allows you to select the state that the analyzer should enter
while waiting for the next analysis to begin if cycle time is greater than the time required for
cell flush and oxidation. You may select without conductivity, with conductivity (default), or
none.
•
If without conductivity is selected, the sample valve opens during idle, allowing
continuous flow through the cell. No conductivity is reported.
•
If with conductivity is selected, the sample valve opens during idle, allowing continuous
flow through the cell. Conductivity of the water is reported.
•
If none, the analyzer sample valve closes, stopping all flow through the sample cell
during idle.
Return icon
returns you to the setup dialog box and saves the changes. Any changes are
noted in the audit trail. All changes take effect on the next change of state.