Specifications
Specifications are subject to change without notice.
Refrigerated Sampler and All Weather Refrigerated Sampler (AWRS)
Specification
Refrigerated Sampler
AWRS
Dimensions (W x D x
H)
1
61 x 61 x 112 cm (24 x 24 x 44 in.)
76 x 81 x 130 cm (30 x 32 x 51 in.)
Weight
63.3 kg (140 lb)
86 kg (190 lb)
Power requirements
(includes compressor)
115 VAC, 60 Hz, 3.3 A (18 locked
rotor amps)
230 VAC, 50 Hz, 1.7 A (9 locked rotor
amps)
115 VAC, 60 Hz, 4.2 A or 6.4 A with controller
compartment heater
230 VAC, 50 Hz, 2.7 A or 4.1 A with controller
compartment heater
Overload protection
Thermal overload relay, opens at
100 to 110 °C (212 to 230 °F)
115 VAC: 7.5 A circuit breaker
230 VAC: 5.0 A circuit breaker
Compressor
1
/
6
HP
1
/
5
HP
115 VAC: 115 °C (239 °F) thermal overload
protector, 7.1 locked rotor amps
230 VAC: 120 °C (248 °F) thermal overload
protector, 7.6 A peak start current
Operating temperature
0 to 50 °C (32 to 122 °F)
0 to 50 °C (32 to 122 °F); with AC battery
backup: 0 to 40 ºC (32 to 104 ºF); with
controller compartment heater: –40 to 50 °C
(–40 to 122 °F); with controller compartment
heater and AC battery backup: –15 to 40 ºC
(5 to 104 ºF)
Storage temperature
–30 to 60 °C (–22 to 140 °F)
–30 to 60 °C (–22 to 140 °F)
Relative humidity
0 to 95%
0 to 95%
Installation category,
pollution degree
II, 2
II, 2
Protection class
I
I
Temperature control
4 (±0.8) °C (39 (±1.5) °F) in ambient
temperatures at a maximum of 50 °C
(120 °F)
4 (±0.8) °C (39 (±1.5) °F)
2
1
on page 6 for the sampler dimensions.
2
Radio frequency interference in the 30 to 50 MHz range can cause a maximum temperature
change of 1.3 °C (34.3 °F). Adjust the set point temperature to 2 to 10 °C (35.6 to 50 °F) to
correct this interference.
English
3
Summary of Contents for AS950 AWRS
Page 2: ...English 3 Fran ais 27 Espa ol 52 Portugu s 77 102 124 2...
Page 22: ...22 English...
Page 47: ...Fran ais 47...
Page 72: ...72 Espa ol...
Page 97: ...Portugu s 97...
Page 105: ...1 AWRS 105...
Page 106: ...106...
Page 108: ...2 AWRS 1 6 11 2 7 12 3 8 13 4 9 AWRS 14 5 10 86 kg 190 lb 90 kg 198 lb 3 108...
Page 110: ...4 AWRS 1 2 2x 111 5 5 110...
Page 111: ...1 6 8 2 8 6 1 24 1 L 4 8 24 7 2 24 350 mL 5 8 AWRS 3 6 9 1 7 PE 2 8 PE 111...
Page 112: ...3 9 102 7 8 112...
Page 113: ...9 1 2 3 GFCI GFI 15 A AWRS 5 10 113...
Page 116: ...pH 12 12 5 POWER POWER AWRS 116...
Page 117: ...13 1 13 2 3 4 5 O 6 13 1 3 O 2 4 117...
Page 118: ...4 6 m 15 2 m 15 ft 50 ft 1 2 14 3 4 5 14 1 7 2 8 AWRS 3 9 4 10 5 11 AWRS 6 12 AWRS 118...
Page 119: ...119...
Page 120: ...118 120...
Page 122: ...16 16 2 2 2071 4652 118 122...
Page 123: ...2 2071 4652 123...
Page 127: ...1 AWRS Web 127...
Page 128: ...128...
Page 130: ...2 AWRS 2 AWRS 2 AWRS 1 6 11 2 7 12 3 8 13 4 9 AWRS 14 5 10 86 Kg 90 kg 3 HACH Japan 130...
Page 131: ...3 1 5 AWRS 2 6 3 7 4 8 3 2000 m 2 000 m AWRS 124 1 127 AWRS 3 8 4 1 2 14 NPT 3 131...
Page 132: ...4 AWRS 1 2 2 1 133 5 5 132...
Page 133: ...1 6 8 1 2 8 6 1 24 x 1 L 4 8 24 7 2 24 x 350 mL 5 8 AWRS 3 1 6 9 133...
Page 134: ...1 7 PE 2 8 PE 3 9 124 7 134...
Page 135: ...8 9 1 2 3 GFCI GFI 135...
Page 139: ...1 4 6 90 18 V 2 200 mA DC 18 VDC 7 RF RF AC AUX I O 3 m 7 12 12 5 POWER POWER AWRS 2 139...
Page 140: ...13 1 13 2 3 4 140...
Page 141: ...5 O 6 13 1 3 O 2 4 4 6 m 15 2 m 1 2 14 3 4 5 141...
Page 142: ...14 1 7 2 8 AWRS 3 9 4 10 5 11 AWRS 6 12 AWRS 142...
Page 143: ...143...
Page 144: ...141 144...
Page 145: ...1 2 4 8 mm 19 mm 15 1 3 4 15 1 4 152 4 mm 177 8 mm 190 8 mm 7 2 5 3 6 145...
Page 146: ...16 16 2 2 AC 146...
Page 147: ...2 2071 4652 141 2071 4652 147...
Page 148: ......
Page 149: ......