22
Installation
Figure 12 SS7 sc-HST plumbing diagram
1
Optional items
14
Sample unit
2
Bubble trap
15
sc100
3
3-way ball valve (Auto Flush Kit)
16
Customer supplied power on/off switch box (NEMA 4X)
required for agency compliance
4
Cooling water to drain
17
Power in for sc100
5
Cooling water out
18
¾-in. NPT adapter (supplied)
6
Flow control valve
19
Drain Trap (Customer-supplied)
7
Sample in
20
1-in. NPT adapter (supplied)
8
Sample bypass during flush cycle
21
To drain
9
Cooling water in
22
¼-in. air purge fitting (50 SCFH instrument air max)
10
Heat exchanger
23
Ball valve (supplied)
11
Electrical box connection
24
¾-in. NPT nipple (supplied)
12
127 mm (5 in.) minimum
25
Customer supplied hose to drain
13
Customer supplied air for flow multiplier
A
Sample during normal operation
D
Cooling water during auto flush
B
Sample bypass during auto flush
E
Electrical
C
Cooling water in normal operation
F
Drain