Arsenic

DOC316.53.01005

Silver Diethyldithiocarbamate Method

1

Method 8013

0 to 0.200 mg/L As

Scope and application: 

For water, wastewater and seawater. Distillation is required. USEPA accepted

2

 for

reporting for drinking and wastewater analysis.

1

Adapted from 

Standard Methods for the Examination of Water and Wastewater

.

2

Procedure is equivalent to Standard Method 3500-As for drinking water analysis.

Test preparation

Instrument-specific information

Table 1

 shows all of the instruments that can make a user calibration program for this

test. The table also shows sample cell and orientation requirements for reagent addition
tests, such as powder pillow or bulk reagent tests.
To use the table, select an instrument, then read across to find the applicable information
for this test.

Table 1  Instrument-specific information

Instrument

Sample cell orientation

Sample cell

DR 6000 
DR 3800 
DR 2800 
DR 2700 
DR 1900 

The fill line is to the right.

2612602

DR 5000 
DR 3900 

The fill line is toward the user.

Before starting

A manual calibration is necessary for this procedure, refer to 

Calibration

 on page 5. Make a new calibration and update

the user program for each new lot of arsenic absorber solution.

Prepare the arsenic absorber solution before starting the test procedure. Refer to 

Reagent preparation

 on page 5.

Set up the distillation apparatus in a fume hood to prevent exposure to hazardous gas.

In bright light conditions (e.g., direct sunlight), close the cell compartment, if applicable, with the protective cover during
measurements.

Do not use the Pour-Thru Cell or sipper module (for applicable instruments) with this test.

Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective
equipment.

Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal
information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory
agencies for further disposal information.

Items to collect

Description

Quantity

Apparatus (refer to 

Required apparatus

 on page 7)

Arsenic Standard Solution, 1000–mg/L As

varies

Hydrochloric Acid, ACS

25 mL

1

Summary of Contents for DR 6000

Page 1: ...Before starting A manual calibration is necessary for this procedure refer to Calibration on page 5 Make a new calibration and update the user program for each new lot of arsenic absorber solution Pre...

Page 2: ...p the preserved samples at room temperature for a maximum of 6 months Before analysis adjust the pH to approximately 7 with 5 N sodium hydroxide solution Correct the test result for the dilution cause...

Page 3: ...ric Acid ACS to the distillation flask 10 Use a serological pipet to add 1 mL of Stannous Chloride Solution to the distillation flask 11 Use a serological pipet to add 3 mL of Potassium Iodide Solutio...

Page 4: ...n the blank sample cell 22 Insert the blank into the cell holder Zero 23 Push ZERO The display shows the non zero intercept that is calculated from the user entered calibration curve 24 Prepare the sa...

Page 5: ...s follows a Use a pipet to add 10 0 mL of the 1000 mg L Arsenic Standard Solution into a 1000 mL volumetric flask b Dilute to the mark with deionized water 2 Prepare three calibration standard solutio...

Page 6: ...FORCE ZERO is off 12 If the graph is acceptable select DONE EXIT 13 Save store the user program When the user program is selected in the test procedure the calibration curve is used to measure the sa...

Page 7: ...s stopper 1 each 1457453 Flask volumetric Class A 200 mL 1 each 1457445 Flask volumetric Class A 500 mL 6 each 1457449 Pipet filler safety bulb 1 each 1465100 Pipet serological 5 mL 1 each 53237 Pipet...

Page 8: ...ANCE PRICE INFORMATION AND ORDERING Call 800 227 4224 Contact the HACH office or distributor serving you www hach com techhelp hach com In the U S A Outside the U S A On the Worldwide Web E mail toll...

Reviews: