background image

Items to collect (continued)

Description

Quantity

Lead Acetate Solution, 10%

1 mL

Potassium Iodide Solution, 20%

3 mL

Pyridine, ACS

50 mL

Sample cells (For information about sample cells, adapters or light shields, refer to 

Instrument-

specific information

 on page 1.)

Silver Diethyldithiocarbamate

1 g

Stannous Chloride Solution

1 mL

Water, deionized

varies

Zinc, 20-mesh, ACS

6 g

Refer to 

Consumables and replacement items

 on page 6 for order information.

Sample collection and storage

Collect samples in clean glass or plastic bottles that have been cleaned with 6 N (1:1)
hydrochloric acid and rinsed with deionized water. 

To preserve samples for later analysis, adjust the sample pH to less than 2 with
concentrated sulfuric acid (approximately 2 mL per liter). No acid addition is
necessary if the sample is tested immediately.

Keep the preserved samples at room temperature for a maximum of 6 months.

Before analysis, adjust the pH to approximately 7 with 5 N sodium hydroxide solution.

Correct the test result for the dilution caused by the volume additions.

Test procedure

W A R N I N G

 

Gas inhalation hazard. Operate the instrument in a fume hood to prevent exposure to
hazardous gas.

User Programs

1. 

 Prepare a calibration and

user program for this
method. Refer to 

Enter the

calibration as a user
program

 on page 5.

Record the program
number.
Select User Programs.

Start

2. 

 Start the user-entered

program from step 1. For
information about sample
cells, adapters or light
shields, refer to 

Instrument-

specific information

on page 1.

3. 

 Prepare the distillation

apparatus, refer to the

Distillation Manual

 for

assembly instructions. 

Do

not connect to the
aspirator.

4. 

 Soak a cotton ball with

10% Lead Acetate Solution.
Put the cotton ball in the gas
scrubber. Make sure that
the cotton seals against the
glass.

2

Arsenic Silver Diethyldithiocarbamate Method (0.200 mg/L)

Summary of Contents for DR 6000

Page 1: ...Before starting A manual calibration is necessary for this procedure refer to Calibration on page 5 Make a new calibration and update the user program for each new lot of arsenic absorber solution Pre...

Page 2: ...p the preserved samples at room temperature for a maximum of 6 months Before analysis adjust the pH to approximately 7 with 5 N sodium hydroxide solution Correct the test result for the dilution cause...

Page 3: ...ric Acid ACS to the distillation flask 10 Use a serological pipet to add 1 mL of Stannous Chloride Solution to the distillation flask 11 Use a serological pipet to add 3 mL of Potassium Iodide Solutio...

Page 4: ...n the blank sample cell 22 Insert the blank into the cell holder Zero 23 Push ZERO The display shows the non zero intercept that is calculated from the user entered calibration curve 24 Prepare the sa...

Page 5: ...s follows a Use a pipet to add 10 0 mL of the 1000 mg L Arsenic Standard Solution into a 1000 mL volumetric flask b Dilute to the mark with deionized water 2 Prepare three calibration standard solutio...

Page 6: ...FORCE ZERO is off 12 If the graph is acceptable select DONE EXIT 13 Save store the user program When the user program is selected in the test procedure the calibration curve is used to measure the sa...

Page 7: ...s stopper 1 each 1457453 Flask volumetric Class A 200 mL 1 each 1457445 Flask volumetric Class A 500 mL 6 each 1457449 Pipet filler safety bulb 1 each 1465100 Pipet serological 5 mL 1 each 53237 Pipet...

Page 8: ...ANCE PRICE INFORMATION AND ORDERING Call 800 227 4224 Contact the HACH office or distributor serving you www hach com techhelp hach com In the U S A Outside the U S A On the Worldwide Web E mail toll...

Reviews: