auxiliary port on the sampler using a full auxiliary cable. If it is necessary to externally power the
logger while connected to an AS950 sampler, use the power input aux cable assembly. To make a
sampler program, refer to the documentation supplied with the sampler.
User interface
The indicators on the user interface show the status of the instrument and the modem. Refer to
Figure 8 User interface
1
Instrument without modem
2
Instrument with modem
Table 3 LED status indicators
Indicator
LED color Description
Green
Flashes every 3 seconds during normal operation. Flashes every
15 seconds during sleep mode.
Red
Flashes when an attached sensor does not agree with the logger
program, an expected sensor is not found or the sensor operation has
failed.
Green
Stays green during a call to the server.
Red
Flashes red if the call to the server failed.
Program a unit equipped with a modem
C A U T I O N
Electromagnetic radiation hazard. Make sure that the antenna is kept at a minimum distance of 20 cm
(7.9 in.) from all personnel in normal use. The antenna cannot be co-located or operated in conjunction
with any other antenna or transmitters.
For units without a modem, refer to
on page 20. For units that come with activated
modem accounts
2
, go to
Add the logger to the FSDATA Server
on page 18.
Set up the wireless account
Note: Adequate cellular coverage from the selected carrier must be verified for each site before a logger with a
cellular modem is purchased.
2
The units that come with activated modem accounts have part numbers with .AX, .AR, .VX,
or .VR as the last digits.
14
English