Option
Description
Stability criteria
When Auto-stabilization is set to No, sets the stabilization rate that
determines when measurements are stable. Options: 0.1 mV/minute to
9.9 mV/minute (default: 1.0 mV/minute). A lower value gives more time for
the measurements to become stable.
Maximum
stability
Sets the maximum time that the meter waits for a measurement to become
stable. Options: 30 to 1200 seconds (default: 600 seconds).
4.
Select Calibration to change the settings for calibrations.
Option
Description
Standard set
Selects the standard solutions to use for calibration from a list of pre-set and
user-specified standard solution values. The set must have a minimum of two
and a maximum of five standards.
Form
Sets the chemical form (applicable to ammonia, ammonium and nitrate
probes).
Options for ammonia probe: NH
3
or NH
3
—N.
Options for ammonium probe: NH
4
+
, NH
4
+
—N or NH
4
+
–NH
3
(ammonium as
ammonia).
Options for nitrate probe: NO
3
or NO
3
—N.
Units
Sets the calibration units. Options: mg/L (default), µg/L, g/L, g/kg, mol/L,
mmol/L, mol/kg, %, ppm or ppb.
Add a standard
Lets the user enter a user-specified standard solution value.
Delete
standards
Removes a user-specified standard solution from the list of available
standard solutions.
Slope limit
Sets the accepted range of values (from theoretical) for the calibration slope.
Options: 1 to 30% (default: 15%).
Reminder
Sets a time interval for probe calibration. A reminder shows on the meter
after the time interval. Options: Off (default), 2 hours, 4 hours, 8 hours,
2 days, 5 days or 7 days. The meter can also make a sound during the
reminder (Settings > Sound).
Expiration
When Reminder is set to a time interval, changes the calibration status to
expired after the reminder occurs. Options: Immediately after reminder,
30 minutes after reminder (default), 1 hour after reminder, 2 hours after
reminder or Continue reading (calibration status does not change).
5.
Select Verification to change the settings for calibration verifications.
Option
Description
Standard
Selects the standard solution to use for calibration verification from a list of
pre-set and user-specified standard solution values.
Form
Sets the chemical form (applicable to ammonia, ammonium and nitrate
probes).
Options for ammonia probe: NH
3
or NH
3
—N.
Options for ammonium probe: NH
4
+
, NH
4
+
—N or NH
4
+
–NH
3
(ammonium
as ammonia).
Options for nitrate probe: NO
3
or NO
3
—N.
Units
Sets the verification units. Options: mg/L, µg/L, g/L, g/kg, mol/L, mmol/L,
mol/kg, %, ppm, ppb.
Add a standard
Lets the user enter a user-specified standard solution value.
28
English