Option
Description
Reminder
Sets a time interval for verification. A reminder shows on the meter after
the time interval. Options: Off (default), 30 minutes, 2 hours, 4 hours,
8 hours, 12 hours or 24 hours.
Snooze
reminders
When Reminder is set to a time interval, lets the user ignore a verification
reminder for a selected time. Options: Yes (default) or No.
Snooze time
When Snooze reminders is set to Yes, sets the time when the verification
reminder will show again. Options: 5 minutes, 10 minutes, 15 minutes
(default), 30 minutes or 60 minutes.
6.
Select Probe settings to manage the probe settings as a group.
Option
Description
Current
Selects the probe settings to use for measurements, calibrations and
verifications. Options: Default, Edited or User-specified name. The Edited
option shows when the current settings have changes but are not saved with
a user-specified name.
Create new
settings
Saves the current probe settings with a user-specified name (maximum
16 characters, letters and numbers).
Delete settings
Removes a user-specified probe setting that is not in use.
9.2 Change the ORP settings
Refer to the steps that follow to change the settings for measurements, calibrations and verifications
with an ORP probe. The user can save the settings with a user-specified name.
1.
Push
and select Settings. The list of settings shows.
2.
Select MTC[probe number].
3.
Select Measurement to change the settings for sample measurements.
Option
Description
Response time
Sets the stabilization time. A slow speed gives more time for the
measurement to become stable. Options: Fast (2 mV/minute), Normal
(1 mV/minute, default) or Slow (0.5 mV/minute).
Lower limit
Sets the minimum accepted value for the measurement.
Options: –1200.0 to 1199.9 mV (default: –1200 mV).
Upper limit
Sets the maximum accepted value for the measurement.
Options: –1199.9 to 1200.0 mV (default: 1200 mV).
Maximum
stability
Sets the maximum time that the meter waits for a measurement to become
stable. Options: 30 to 600 seconds (default: 300 seconds).
4.
Select Calibration to change the settings for calibrations.
Option
Description
Standard
Sets the standard solution to use for calibration. Options: Zobell's (221 mV -
25 °C) (default), Light's (468 mV - 25 °C) or Custom (user-specified mV value).
Measurements are temperature compensated when Zobell's solution is used
for calibration.
Standard
value
WhenStandard is set to Custom, sets the mV value for custom standard
solutions. Options: –1200 mV to 1200 mV (default: 221 mV).
English
21