Table 1 Spreadsheet column descriptions (continued)
Column header name
Data description and example values
Any settings that affect the
reading, for example
“NaCl/Non-Linear”
Reading Message 1–4 Any message that was shown during the measurement,
for example “Out of limits”.
Check Std Value
Value of the check standard that was used to verify accuracy, for example:
7.00 pH–25 ºC (pH, temp-compensated); 7.01 pH (pH, custom)
Check Std Units
Check standard units, for example μS/cm.
Note: pH is not displayed here as it is included in the previous column.
Check Std Graph
Bar-graph showing the measurement in relation to the acceptance limits.
Example: “6.901 <—|—> 7.101”.
Check Std Status
Status of the check standard reading. Example: “Reading within limits”,
“Reading outside limits”
Calibration Status
Status of the calibration that is in use.
= current calibration is valid;
= calibration has expired.
Cal Date
Date of calibration reading: stored in user-defined date format
Cal Time
Time of calibration reading: stored in user-defined time format
Cal Operator ID
The operator ID specified when the probe was calibrated. Shows “- - -” if
undefined.
Cal Slope Name
Slope (pH or LDO) or cell constant (conductivity)
Cal Slope
The slope value for the calibration
Cal Slope Aux
Used by pH to give the percent of nominal slope
Cal Slope Units
Units of the calibration slope. Example: “mV/pH” for pH
Cal Offset
Calibration offset value
Cal Offset Units
Calibration offset units. Example: “mV” for pH.
Cal r²
Calibration correlation coefficient without a unit (may be blank)
Cal Number of Std’s
Number of standards used during calibration, for example 5. May be blank
depending on record type, parameter type, and method settings.
Cal Std 1
Known value of the first calibration standard
Cal Std 1 Units
Units of the first calibration standard
Cal Std 1 Primary Value
Measured value of the first calibration standard
Cal Std 1 Primary Units
Associated units for the calibration measurement
Cal Std 1 Supp Value
Value of supplemental measurement, for example temperature
Cal Std 2–7
Known value of additional calibration standards, if used
Cal Std 2–7 Units
Units of additional calibration standards, if used
Cal Std 2–7 Primary Value
Measured values of additional calibration standards, if used
Cal Std 2–7 Primary Units
Associated units for additional calibration measurements, if used
Cal Std 2–7 Supp Value
Value of supplemental measurement, for example temperature
Cal Std Supp Units
Units applicable to all secondary calibration readings. Example: “ºC” or “ºF” for
temperature
Cal Message 1–4
Any messages about the calibration