background image

Download software updates

To download the most current version of the software, refer to the applicable product page on the
manufacturer's website.

1.

Transfer the update files to a USB storage device.

2.

Follow the instructions in 

Update the meter software

 on page 22 to update the software in the

meter.

Transfer method settings

Probe settings that have been changed by the user for measurements, calibrations or check
standards (Meter Options > (Probe Model) Settings > Modify Current Settings) can be copied to a
USB storage device. The USB device can then be used to transfer the method settings to other HQd
meters that accept the same probes.

N O T I C E

 

Make sure the USB storage device does not contain HQd meter software update files to prevent unintentional
updates.

1.

Turn off the meter.

2.

Connect the meter to AC power (

Figure 3

 on page 6).

3.

Insert the USB storage device.

4.

Turn on the meter.

5.

Push the 

 key and select Transfer Methods. If the USB device already contains a method

settings file, an option to export or import methods is shown. Select Export Methods.

6.

In the Select Methods to Export screen, select one or more methods to copy to the USB device. A
check mark is shown next to each selected method.

7.

Select OK. The settings are copied to the USB storage device. When complete, the Transfer
Summary screen is shown.

8.

Connect the AC power and USB device to a meter that will receive the method settings. Turn the
meter on.

9.

Push the 

 key and select Transfer Methods. If the USB device already contains a method

settings file, an option to export or import methods is shown. Select Import Methods.

10.

In the Select Methods to Import screen, select one or more methods to transfer to the meter. A
check mark is shown next to each selected method.

11.

Select OK. The user method settings are transferred from the USB storage device to the meter.
When complete, the Transfer Summary screen is shown. Select details to view additional
information about the transfer.

12.

Disconnect the USB storage device from the meter.

Bi-directional Communication between the meter and a PC

For measurement automation the meter can be used to implement a command set for meter remote
control or automated data transfer. The command set can be used to perform minimal configuration
and to control the meter. To set up the meter for communication and control, refer to 

About meter

configuration

 on page 23. For additional information and the command set contact Technical

Support.

About meter control

The virtual serial connection can be used to control meter functions from a PC. For example, the
functions include starting a measurement cycle, turning off the meter and sending the entire
measurement (including calibrations) to the PC or other information management system.

About meter configuration

To use the meter communication and control from the PC, an INF file must be installed.

English

   

23

Summary of Contents for HQ440d

Page 1: ...DOC022 53 80116 HQ440d HQ430d HQ411d 10 2016 Edition 3 User Manual...

Page 2: ......

Page 3: ...on 11 About calibration 11 About sample measurements 11 About check standards 11 Use a sample ID 12 Use an operator ID 12 Data management 12 About stored data 12 View stored data 12 View stored probe...

Page 4: ...information 22 Update the meter software 22 Download software updates 23 Transfer method settings 23 Bi directional Communication between the meter and a PC 23 About meter control 23 About meter confi...

Page 5: ...2 0 flash memory device printer keyboard and Integrated USB type B for PC Temperature correction Off automatic and manual parameter dependent Measurement display lock Continuous measurement Interval o...

Page 6: ...thout care Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems Return old or end of life equipment to the manufacturer for disposal at n...

Page 7: ...device printer and keyboard 3 USB connector type B for PC connections 6 Probe connection port Product components Refer to Figure 2 to make sure that all components have been received If any items are...

Page 8: ...rrect battery installation can cause the release of explosive gases Be sure that the batteries are of the same approved chemical type and are inserted in the correct orientation Do not mix new and use...

Page 9: ...ace the battery cover Note Do not over tighten the screws Figure 4 Battery installation User interface and navigation User interface Figure 5 Keypad description 1 LEFT key calibrates cancels or exits...

Page 10: ...atus 12 Display size icon 5 Power source 13 Calibrate Cancel Exit 6 Sample temperature C or F 14 Sample and operator identification 7 Secondary measurement unit 15 Stability or display lock indicator...

Page 11: ...Figure 8 Note For probe calibration change the screen mode to the single screen mode To change the screen mode to single or dual screen use the and keys In dual screen mode the key will select the pr...

Page 12: ...ged from the Meter Options menu Access to the language menu can be restricted with the Security Options Refer to Security options on page 17 1 Push the key and select Language 2 Select a language from...

Page 13: ...red the LDO sensor cap should be replaced the calibration is out of range the calibration results are outside acceptance criteria settings About sample measurements Each probe has specific preparation...

Page 14: ...ed Delete Operator ID Delete an existing operator ID Data management About stored data The following types of data are stored in the data log Sample measurements stored automatically each time a sampl...

Page 15: ...rent calibration information shows the calibration details for the most recent calibration If the probe has not been calibrated by the user the factory calibration data is shown View Calibration Histo...

Page 16: ...contain the same information as in the advanced report Send data to a USB storage device N O T I C E The transfer of a large number of data points will take some time DO NOT disconnect the USB storage...

Page 17: ...IC Current Calibration Parameter Type Parameter LDO pH CD conductivity ORP ISE Date Date of reading stored in user defined date format Time Time of reading stored in user defined format Operator ID O...

Page 18: ...e value for the calibration Cal Slope Aux Used by pH to give the percent of nominal slope Cal Slope Units Units of the calibration slope Example mV pH for pH Cal Offset Calibration offset value Cal Of...

Page 19: ...the meter is connected to AC power 3 Connect the PC to the meter with a USB type B cable 4 Turn on the meter 5 Open the HQ40d PC Application on the computer Click on the green triangle in the menu ba...

Page 20: ...to enable the Full Access Options menu again when the meter is turned on Set Password Set a password that must be entered to enable the Full Access Options menu The requirement for password entry is...

Page 21: ...connected to use this option Instrument information Probe information Meter information Access password Enter password Display options Contrast Auto shutoff Backlight Mode Sounds Key press Stability a...

Page 22: ...is reached Cal reminder The meter will make an audible sound when calibration is due Note Refer to Set the probe calibration reminder on page 20 to set the calibration reminder to on or off Set the p...

Page 23: ...h 8 h 24 h 48 h or no limit 5 Select Interval and select how often measurements will be taken every 10 s 30 s 1 min 5 min 15 min or 30 min Start interval measurements During interval measurements the...

Page 24: ...then on again before the software update will begin The software update initiates upon meter startup after the USB device is correctly inserted N O T I C E Do not remove the USB device until the Upda...

Page 25: ...he AC power and USB device to a meter that will receive the method settings Turn the meter on 9 Push the key and select Transfer Methods If the USB device already contains a method settings file an op...

Page 26: ...on in the search and click the Browse button 15 Select the INF folder or location and click OK 16 Click Next The new software will be installed 17 Click Finish to complete the Found New Hardware Wizar...

Page 27: ...ery replacement Troubleshooting Refer to the following table for common problem messages or symptoms possible causes and corrective actions Error Warning Description Solution Connect a Probe Probe dis...

Page 28: ...meter batteries 3 Install the meter batteries properly Follow the polarity makings 4 Set correct date and time in the meter 5 Connect the probe and verify that message has been removed Software not u...

Page 29: ...B male 5924000 Power supply without power cord 5827000 Power cord US 1801000 Power cord EU YAA080 Power cord UK XLH057 Power cord CH XLH051 Keyboard QWERTY USB type LZV582 USB memory device with HQd S...

Page 30: ...n a basic advanced or total report format Total report refer to Figure 13 Advanced report refer to Figure 14 Basic report refer to Figure 15 Figure 13 Basic report for sample data 1 line 1 Data type R...

Page 31: ...ype of parameter and user settings May be blank 3 Date of calibration DD MM YY or user defined format 6 Operator ID user defined shows if undefined Check standard reports Check standard data is printe...

Page 32: ...d and number of calibration standards used May be blank 5 Probe model 11 Operator ID user defined shows if undefined 6 Date of calibration DD MM YY or user defined format The calibration history can b...

Page 33: ...ay description 19 dual screen mode calibration 9 I interval measurements 20 M measurement modes 20 menu navigation 9 modes screen 19 O operator ID 12 P power 6 battery replacement 25 printing 13 probe...

Page 34: ......

Page 35: ......

Page 36: ...2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 ro...

Reviews: