Option
Description
06
Sets the month and day (MM-DD).
07
Sets the year (YY).
3.
Push
ENTER
.
4.
Use the arrow keys to change the value.
5.
Push
ENTER
.
6.
Push
SETUP
.
Show the current time (optional)
1.
Push
SETUP
. The SETUP light turns on.
2.
Use the arrow keys to select 08.
3.
Push
ENTER
. The current time is shown on the display (HH-MM-
SS).
4.
Push
SETUP
.
Standard operation
Calibrate the turbidimeter with StablCal
®
Standards
Calibrate the turbidimeter before it is used for the first time using the
StablCal
®
sealed vial standards provided. As an alternative, calibration
can be done with recently prepared formazin standards.
Calibrate the turbidimeter at least every 3 months or as specified by the
regulating authority when data is used for USEPA reporting.
The instrument is ready for calibration 60 minutes after start-up. Keep
the instrument on 24 hours a day if the instrument is used regularly.
Note: Unknown results may occur if standards other than the recommended
calibration points are used. The recommended calibration points (< 0.1, 20, 200,
1000, 4000 and 7500 NTU) provide the best calibration accuracy. Use of standards
other than StablCal, or user-prepared formazin, may result in less accurate
calibrations. The manufacturer cannot guarantee the performance of the instrument
if calibrated with co-polymer styrenedivinylbenzene beads or other suspensions.
Prepare the StablCal standards
When received and at intervals:
1.
Clean the exterior surface of the StablCal vials with laboratory glass
cleaning detergent.
2.
Rinse the vials with distilled or deionized water.
3.
Dry the vials with a lint-free cloth.
Note: Never shake or invert the < 0.1 NTU standard. If the standard has been
mixed or shaken, do not move the vial for 15 minutes or more before using.
Note: Do not remove the caps from the sealed vials.
Make sure that the StablCal standards are at ambient instrument
temperature before use (and no greater than 40 °C (104 °F)).
Mix the standards before use:
1.
Open the case lid. Remove the < 0.1 NTU standard from the plastic
case.
2.
Leave the other standards in the case. Close the case lid.
3.
Shake the case vigorously for at least 10 seconds.
4.
Let the standards stand with no movement for 3–5 minutes before
use.
Calibration notes
• Make sure that the instrument is in the same ambient conditions as
where it is used.
• Make sure that the standards are at the same ambient temperature as
the instrument before use.
• Use only the provided silicone oil. This silicone oil has the same
refractive index as the vial glass and masks minor glass differences
and scratches.
• Store the oiling cloth in a plastic storage bag to keep the cloth clean.
• If power is lost during calibration, the new calibration data is lost and
the last calibration data is used. To exit a calibration and not save the
new values, push
UNITS/Exit
.
• In Calibration mode, automatic range and signal averaging on are
selected. When calibration is completed, all operational modes go
back to the last settings.
• All nephelometric (turbidity units of measure) calibrations are done at
the same time.
English
11
Summary of Contents for Lange 2100AN
Page 18: ...1 18 English...
Page 39: ...1 Deutsch 39...
Page 60: ...1 60 Italiano...
Page 80: ...1 80 Fran ais...
Page 102: ...1 102 Espa ol...
Page 122: ...1 122 e tina...
Page 142: ...1 142 Dansk...
Page 162: ...1 162 Nederlands...
Page 182: ...1 182 Polski...
Page 202: ...1 202 Svenska...
Page 222: ...1 222 Suomi...
Page 241: ...1 Magyar 241...
Page 261: ...1 Rom n 261...
Page 281: ...1 T rk e 281...
Page 288: ...12 2005 2002 96 IECS 003 A IECS IECS FCC 15 A 15 FCC 1 2 15 FCC 1 2 3 4 5 288...
Page 292: ...1 Calibration Setup 1 NTU NEP EBC CU T Abs 292...
Page 296: ...StablCal 1 302 2 USEPA 302 3 4 CAL Zero CAL Zero 00 NTU 5 0 1 NTU 6 296...
Page 298: ...30 60 24 298...
Page 299: ...1 30 mL 2 3 4 5 6 299...
Page 300: ...7 RS232 PRINT RANGE MANUAL AUTO 40 NTU 268 nephelo 9 8 EBC 300...
Page 302: ...1 2 3 1 2 3 302 4 5 1 2 3 30 60 302...
Page 303: ...1 303...
Page 307: ......