1.
Pull the tab on the docking module away from the instrument to remove
the module.
2.
Remove the four screws on the bottom plate of the module. Separate
the bottom plate from the module enclosure.
3.
Pull to remove the stirrer plug from the circuit board.
4.
Remove the two holding screws from the stirrer motor flange and
remove the magnetic stirrer assembly.
5.
Put the bottom plate on the module enclosure and install the four
screws.
User interface
The buttons on the front panel allow the user to operate the syringe,
sample probe and stirrer. Refer to
,
Figure 6 Front panel
1
Syringe controls
4
Stop button
2
Stirrer speed (RPM)
5
Sample probe controls
3
Power switch
6
LED status indicators
Table 1 Button descriptions
Button Name
Description
Load
Sends the syringe plunger to the top or bottom position.
Clean
Flushes the system.
Back-flush Back-flushes the system when held for 3 seconds.
14
English
Summary of Contents for LANGE HIAC 9703+
Page 2: ...English 3 Deutsch 20 Italiano 37 Fran ais 54 Espa ol 71 2...
Page 9: ...1 2 3 4 English 9...
Page 10: ...5 7 6 10 English...
Page 11: ...1 2 3 4 English 11...
Page 12: ...5 7 6 12 English...
Page 26: ...1 2 3 4 26 Deutsch...
Page 27: ...5 7 6 Deutsch 27...
Page 28: ...1 2 3 4 28 Deutsch...
Page 29: ...5 7 6 Deutsch 29...
Page 43: ...1 2 3 4 Italiano 43...
Page 44: ...5 7 6 44 Italiano...
Page 45: ...1 2 3 4 Italiano 45...
Page 46: ...5 7 6 46 Italiano...
Page 60: ...1 2 3 4 60 Fran ais...
Page 61: ...5 7 6 Fran ais 61...
Page 62: ...1 2 3 4 62 Fran ais...
Page 63: ...5 7 6 Fran ais 63...
Page 77: ...1 2 3 4 Espa ol 77...
Page 78: ...5 7 6 78 Espa ol...
Page 79: ...1 2 3 4 Espa ol 79...
Page 80: ...5 7 6 80 Espa ol...
Page 88: ...www hach lange com Hach Lange GmbH 2010 All rights reserved Printed in Germany...