Figure 7 Attach the syringe
1
Plunger bolt
3
Syringe drive
2
Syringe
4
Valve
1.
Push the
ON/OFF
switch to turn the instrument on.
2.
Push the
LOAD
button under the syringe compartment to set the
plunger bolt at the lowest point. Refer to
3.
Loosen and remove the plunger bolt from the syringe drive.
4.
Hold the syringe and push the bolt through the plunger hole.
5.
Push the bolt in the syringe drive hole and turn the bolt by hand to
tighten.
6.
Pull the syringe barrel up and install the syringe threads in the valve.
7.
Turn the syringe by hand to tighten.
Note: Do not overtighten.
Set the sample probe position
The lift mechanism for the sample probe can be set to automatically go to
the correct position for sampling.
16
English
Summary of Contents for LANGE HIAC 9703+
Page 2: ...English 3 Deutsch 20 Italiano 37 Fran ais 54 Espa ol 71 2...
Page 9: ...1 2 3 4 English 9...
Page 10: ...5 7 6 10 English...
Page 11: ...1 2 3 4 English 11...
Page 12: ...5 7 6 12 English...
Page 26: ...1 2 3 4 26 Deutsch...
Page 27: ...5 7 6 Deutsch 27...
Page 28: ...1 2 3 4 28 Deutsch...
Page 29: ...5 7 6 Deutsch 29...
Page 43: ...1 2 3 4 Italiano 43...
Page 44: ...5 7 6 44 Italiano...
Page 45: ...1 2 3 4 Italiano 45...
Page 46: ...5 7 6 46 Italiano...
Page 60: ...1 2 3 4 60 Fran ais...
Page 61: ...5 7 6 Fran ais 61...
Page 62: ...1 2 3 4 62 Fran ais...
Page 63: ...5 7 6 Fran ais 63...
Page 77: ...1 2 3 4 Espa ol 77...
Page 78: ...5 7 6 78 Espa ol...
Page 79: ...1 2 3 4 Espa ol 79...
Page 80: ...5 7 6 80 Espa ol...
Page 88: ...www hach lange com Hach Lange GmbH 2010 All rights reserved Printed in Germany...