Reference sample
1.
Run the reference sample through the instrument and adjust the flow to approximately
150 mL/min.
2.
In the
Calibration mode
window select
Reference sample
3.
Scroll down to the
Reference value
window and enter the oxygen value of the sample
4.
Let the sample run through the instrument for 5 minutes to stabilize the measurement
5.
Scroll down to the
Start Calibration
window and press the
Enter
key to start the calibration
6.
Wait until the
Signal in range
and
Stability reached
fields display
Yes
which indicates the
calibration is within acceptable limits. The
Calibration possible
field should also display
Yes
at
this point. Press the
Enter
key to complete the calibration
7.
Accept (
OK
) or reject (
Cancel
) the new calibration data
Ref. gas bottle
1.
Rinse flow path with 20 mL of ethanol (EtOH) using the syringe and connectors supplied in the
tool kit
2.
In the
Calibration mode
window select
Ref. gas bottle
3.
Scroll down to the
Reference value
window and enter the oxygen value of the reference gas in
%Vbar
4.
Connect the gas sample to the instrument and adjust the flow to approximately 100 mL/min.
5.
Let the gas mixture run through the instrument for 5 minutes to stabilize the measurement
6.
Scroll down to the
Start Calibration
window and press the
Enter
key to start the calibration
7.
Wait until the
Signal in range
and
Stability reached
fields display
Yes
which indicates the
calibration is within acceptable limits. The
Calibration possible
field should also display
Yes
at
this point. Press the
Enter
key to complete the calibration
8.
Accept (
OK
) or reject (
Cancel
) the new calibration data
Default measurement configuration settings
1.
Instrument settings
Option
Description
Instrument mode
Select between continuous or sample mode.
Continuous mode is typically used for process measurement, whereas sample mode is
aimed at lab measurements of small volume individual samples such as cans, bottles,
etc.
Sample type
Select the sample type from the list available
2.
O2 channel settings
Option
Description
O2 gas unit type
Select the gas unit type from the list available
O2 gas unit
Select the display unit from the list available
English
17
Summary of Contents for Lange ORBISPHERE 3100
Page 247: ...TFT 72 x 54 mm 2 83 x 2 13 ins 1 x USB 5 VDC 1 x RS232 0 5 V 3100 247...
Page 248: ...10 60 C Hach Lange 60 C 10 45 C Hach Lange Hach Lange 248...
Page 249: ...ORBISPHERE 3100 5 760 3 4mm 6mm 1 4 Hach Lange 249...
Page 251: ...4 OFF 1 1 4 USB 2 ON OFF 5 3 RS232 6 2 1 2 PURGE ON OFF PURGE 5 251...
Page 255: ...1 Cancel 5 2 USB 6 RFID 3 7 4 6 4 RFID USB RFID USB USB 10 USB Import Export Main Menu 255...
Page 256: ...Measurement stopped Enter 0000 0 9999 253 254 Enter Supervisor1 5678 256...
Page 257: ...1 7 2 8 USB USB 3 9 4 10 5 11 6 12 5 5 5 4 257...
Page 258: ...254 254 ID0 262 Enter Enter Cancel Enter 258...
Page 259: ...Enter 1 2 12 24 3 4 1 5 760 259...
Page 265: ...1 2 3 4 Nb Reset service timer Last service date 2 1 2 2 2 2 DC 2 DC LED 265...
Page 267: ...1 2 No 1 2 No 2 3 4 5 6 2 7 60 150 200 mL min PURGE 5 2 N2 267...
Page 269: ...No 8 2 251 1 20 mL EtOH N2 No 8 2 251 4 269...
Page 359: ...1 x USB 5 VDC 1 x RS232 0 5 V 3100 359...
Page 360: ...10 60 C Hach Lange 60 C 10 45 C Hach Lange 360...
Page 361: ...Hach Lange ORBISPHERE 3100 5 760 3 4mm 6mm 6 35mm 361...
Page 362: ...Hach Lange 1 362 2 2 1 363 364 3 367 4 371 5 3100 365 6 366 7 366 8 366 1 1 2 2 3 OFF ON 362...
Page 363: ...4 1 1 4 USB 2 5 3 RS232 6 2 1 2 363...
Page 367: ...Enter 6 1 Cancel 5 Enter 2 USB 6 RFID 3 7 4 6 4 RFID USB RFID USB USB 367...
Page 368: ...10 USB Measurement stopped Enter 0000 0 9999 366 366 Enter 368...
Page 369: ...Supervisor1 1 5678 1 7 2 8 USB USB 3 9 4 10 5 11 6 12 5 5 5 369...
Page 370: ...4 y 366 366 ID0 374 Enter Enter Cancel Enter 370...
Page 379: ...8 365 4 DG33228 1 2 1 2 2 3 4 5 6 2 7 379...
Page 380: ...60 150 200 mL min PURGE 5 2 2 LDO Hach Lange N2 CO2 50 mbar 2 LDO 20 mL EtOH 2 5 380...
Page 382: ......
Page 383: ......