6.4 Validate the probe calibration
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and
national regulations.
C A U T I O N
Ultraviolet (UV) light exposure. Exposure to UV light can cause eye and skin damage. Do not look
directly into the measurement path when probe is in operation. Protect eyes and skin from direct
exposure to UV light. Wear all appropriate personal protective equipment.
Do the steps that follow to validate the probe calibration.
1.
Go to the main menu, then select
Sensor setup
.
2.
Select the applicable sensor, if necessary.
3.
Select
Maintenance
.
4.
Set the outputs on
Hold
.
5.
Remove the probe from the tank or flow through unit.
6.
Rinse the measuring path with distilled water.
7.
Prepare the probe as follows:
a.
Fully clean and dry the back hole area and apply tape to cover the hole of the measuring path.
Refer to the illustrated steps in
Note: Make sure that no soluble compounds touch the medium that fills the measuring path.
b.
Turn the probe until the measuring path is in the horizontal position.
c.
Purge the measuring path with 10 mL of standard solution.
d.
Fill the measuring path with standard solution.
8.
Go to the signals menu as follows:
• SC200 and SC1000 Controllers—Go back to the sensor menu, then select
DIAG/TEST
>
Signals
.
• SC4500 Controller and Claros interface—Go back to the sensor menu, then select
Diagnostics/Test
>
Signals
. Select
Enter
to start the measurement.
9.
Examine the values that show on the controller display:
• The first line (first three lines for the NT3200sc) shows the values from the user calibration
(offset, factor, signal average and matrix calibration for the NT3200sc, if applicable).
• The next line (next three lines for the NT3200sc) shows the values from the factory calibration
(factor, offset or matrix calibration).
Remove the tape and make sure that the back hole of the measuring path is free.
10.
Install the probe in the tank or flow through unit.
11.
Select
Back
.
12.
Complete the instructions on the display.
English
21
Summary of Contents for NT3100sc
Page 2: ...Table of Contents English 3 Fran ais 27 Espa ol 52 Portugu s 78 102 124 147 169 2...
Page 11: ...Figure 3 Bracket installation English 11...
Page 20: ...Figure 8 Replace wiper blade 1 mm and 2 mm Figure 9 Replace wiper blade 5 mm 20 English...
Page 86: ...Figura 3 Instala o do suporte 86 Portugu s...
Page 104: ...2 1 2 1 1 2 1 2 104...
Page 108: ...2 6 2 2 1 NT3100sc NT3200sc 3 9 5 mm 5 2 9 1 mm 2 mm 5x 4 LZY261 3 3 1 OFF 3 2 3 5 4 9 108...
Page 109: ...3 109...
Page 110: ...4 5 110...
Page 118: ...9 5 mm 6 4 MSDS SDS UV 1 2 3 4 5 6 7 a 10 b 118...
Page 126: ...2 1 1 2 1 2 126...
Page 130: ...2 6 2 2 1 NT3100sc NT3200sc 3 5 mm 5 9 2 9 1 mm 2 mm 5 4 LZY261 3 3 1 9 130...
Page 131: ...3 2 3 5 4 3 131...
Page 132: ...4 5 132...
Page 133: ...3 3 SC 6 6 4 5 5 1 5 2 1 SC200 2 06 2 SC200 SC1000 SC200 SC4500 Claros 3 133...
Page 140: ...8 1 mm 2 mm 9 5 mm 140...
Page 142: ...10 7 5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 UNSTABLE 142...
Page 143: ...5 6 NO3 NO3 N NO3 NO3 N NO3 NO3 N NO2 NO2 N NO2 NO2 N NO2 NO2 N 8 Web 143...
Page 149: ...2 1 1 2 1 2 UV 149...
Page 153: ...2 6 2 2 1 NT3100sc NT3200sc 3 9 5mm 5 2 9 1mm 2mm 5x 4 LZY261 3 3 1 OFF 3 2 3 5 9 153...
Page 154: ...4 3 154...
Page 155: ...4 5 155...
Page 156: ...3 3 SC 6 6 4 ON 5 5 1 5 2 1 SC200 2 06 2 SC200 SC1000 SC200 SC4500 Claros 3 156...
Page 163: ...9 5mm 6 4 MSDS SDS UV 1 2 3 4 5 6 7 a 10 b 163...
Page 164: ...c 10mL d 8 SC200 SC1000 SC4500 Claros Enter 9 NT3200sc NT3200sc NT3200sc 10 11 12 10 7 5 164...
Page 165: ...5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 6 165...
Page 171: ...5 1 2 PUR HT CE CMIM UKCA FCC ISED 1 2 2 2 1 2 1 1 171...
Page 172: ...2 1 2 UV 2 1 3 2 1 4 ICES 003 Class A Class A 172...
Page 175: ...1 1 3 2 2 6 2 2 1 NT3100sc NT3200sc 3 9 5 5 2 9 1 2 5 4 LZY261 9 175...
Page 176: ...3 3 1 OFF 3 2 3 5 4 176...
Page 177: ...3 177...
Page 178: ...4 5 178...
Page 186: ...9 5 6 4 MSDS SDS UV 1 Sensor setup 2 3 Maintenance 4 Hold 5 6 7 a 10 b 186...
Page 193: ......