Figure 10 Prepare the probe
Section 7 Troubleshooting
Table 5 Error messages
Message
Possible cause
Solution
System error
There is a problem with the
electronic components.
Contact technical support.
Measurement failed
The sample signal is too low.
The temperature is not stable or is
out of range.
The hardware or electronics are
defective.
Contact technical support.
NO3 is too high
The nitrate (NO
3
) concentration is
more than the maximum limit.
Measure the concentration of a grab sample.
Check the measurement path and clean if
necessary.
Calibrate the sensor.
NO3 is too low
The nitrate (NO
3
) concentration is
less than the minimum limit.
NO2 is too high
The nitrite (NO
2
) concentration is
more than the maximum limit.
NO2 is too low
The nitrite (NO
2
) concentration is
less than the minimum limit.
NOx is too high
The nitrate (NOx) concentration is
more than the maximum limit.
NOx is too low
The nitrate (NOx) concentration is
less than the minimum limit.
Unstable
measurement
(sc200/sc1000:
UNSTABLE)
The particles in the medium cause
the measurement to change
continuously.
Check the sensor location and if necessary,
measure with filtration (bypass, flow through
cell).
Matrix calibration
A matrix calibration is necessary.
Complete a matrix calibration. First delete all
existing calibration points when the probe is
moved to a different environment or the matrix
changes significantly.
22
English
Summary of Contents for NT3100sc
Page 2: ...Table of Contents English 3 Fran ais 27 Espa ol 52 Portugu s 78 102 124 147 169 2...
Page 11: ...Figure 3 Bracket installation English 11...
Page 20: ...Figure 8 Replace wiper blade 1 mm and 2 mm Figure 9 Replace wiper blade 5 mm 20 English...
Page 86: ...Figura 3 Instala o do suporte 86 Portugu s...
Page 104: ...2 1 2 1 1 2 1 2 104...
Page 108: ...2 6 2 2 1 NT3100sc NT3200sc 3 9 5 mm 5 2 9 1 mm 2 mm 5x 4 LZY261 3 3 1 OFF 3 2 3 5 4 9 108...
Page 109: ...3 109...
Page 110: ...4 5 110...
Page 118: ...9 5 mm 6 4 MSDS SDS UV 1 2 3 4 5 6 7 a 10 b 118...
Page 126: ...2 1 1 2 1 2 126...
Page 130: ...2 6 2 2 1 NT3100sc NT3200sc 3 5 mm 5 9 2 9 1 mm 2 mm 5 4 LZY261 3 3 1 9 130...
Page 131: ...3 2 3 5 4 3 131...
Page 132: ...4 5 132...
Page 133: ...3 3 SC 6 6 4 5 5 1 5 2 1 SC200 2 06 2 SC200 SC1000 SC200 SC4500 Claros 3 133...
Page 140: ...8 1 mm 2 mm 9 5 mm 140...
Page 142: ...10 7 5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 UNSTABLE 142...
Page 143: ...5 6 NO3 NO3 N NO3 NO3 N NO3 NO3 N NO2 NO2 N NO2 NO2 N NO2 NO2 N 8 Web 143...
Page 149: ...2 1 1 2 1 2 UV 149...
Page 153: ...2 6 2 2 1 NT3100sc NT3200sc 3 9 5mm 5 2 9 1mm 2mm 5x 4 LZY261 3 3 1 OFF 3 2 3 5 9 153...
Page 154: ...4 3 154...
Page 155: ...4 5 155...
Page 156: ...3 3 SC 6 6 4 ON 5 5 1 5 2 1 SC200 2 06 2 SC200 SC1000 SC200 SC4500 Claros 3 156...
Page 163: ...9 5mm 6 4 MSDS SDS UV 1 2 3 4 5 6 7 a 10 b 163...
Page 164: ...c 10mL d 8 SC200 SC1000 SC4500 Claros Enter 9 NT3200sc NT3200sc NT3200sc 10 11 12 10 7 5 164...
Page 165: ...5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 6 165...
Page 171: ...5 1 2 PUR HT CE CMIM UKCA FCC ISED 1 2 2 2 1 2 1 1 171...
Page 172: ...2 1 2 UV 2 1 3 2 1 4 ICES 003 Class A Class A 172...
Page 175: ...1 1 3 2 2 6 2 2 1 NT3100sc NT3200sc 3 9 5 5 2 9 1 2 5 4 LZY261 9 175...
Page 176: ...3 3 1 OFF 3 2 3 5 4 176...
Page 177: ...3 177...
Page 178: ...4 5 178...
Page 186: ...9 5 6 4 MSDS SDS UV 1 Sensor setup 2 3 Maintenance 4 Hold 5 6 7 a 10 b 186...
Page 193: ......