Maintenance and Troubleshooting
8.
Tighten the protection cap finger tight. Then, complete the process using the tool provided
in the maintenance kit. Insert into each of the four holes in turn, and tighten as far as
possible. Tighten each hole only once.
Note:
The grille inside the protection cap should be free to move during tightening. Therefore, and to avoid
damage to the membrane, do not touch the grille during the tightening process.
9.
Always store the sensor in a dry environment, with the storage cap installed.
Note:
A sensor that has been taken apart, or serviced must always be calibrated. Allow the sensor to settle
for 30 minutes to allow measurements to stabilize and the membrane(s) to relax, before performing the
sensor calibration.
Calibrate the sensor to check that the membrane has been installed correctly and not been
damaged. If an error message (illustrated right) appears on your instrument, then the membrane
has been damaged or incorrectly installed.
Figure 49 Installing the membrane
Summary of Contents for Orbisphere 51 Series
Page 5: ...4 Table of Contents...
Page 19: ...18 Specifications...
Page 48: ...47 Section 6 View Menu Figure 33 View menu...
Page 52: ...51 Section 7 Measurement Menu Figure 36 Measurement menu...
Page 66: ...65 Section 9 Inputs Outputs Menu Figure 40 Inputs Outputs menu...
Page 68: ...67 Inputs Outputs Menu 9 4 Analog outputs Figure 41 Analog outputs menu...
Page 75: ...74 Inputs Outputs Menu...
Page 91: ...90 Communication Menu...
Page 94: ...93 Section 12 Products Menu Figure 44 Products menu...
Page 97: ...96 Global Configuration Menu...
Page 98: ...97 Section 14 Services menu Figure 46 Services menu Part 1...
Page 99: ...98 Services menu Figure 47 Services menu Part 2...
Page 103: ...102 Services menu...