12
Specifications
Digital outputs
Measurement alarm relays on the
measurement board(s)
Three alarm relays per measurement board
1A-30 VAC or 0.5A-50 VDC on a resistance load
Configurable to Normally Open [NO] or Normally Closed [NC] contacts by changing the
jumper positions.
WA R N I N G
Potential Electrocution Hazard. Connect only safety low voltage < 33 VAC RMS.
System alarm relay on the main
board. One system board with
one system relay per instrument
One "instrument system alarm" relay per instrument
1A-30 VAC or 0.5A-50 VDC on a resistance load
Normally closed [NC] (NO relay also available) when instrument is turned on. Opens when
a system alarm is detected, and when it does not receive any signal.
WA R N I N G
Potential Electrocution Hazard. Connect only safety low voltage < 33 VAC RMS.
General
Thermal cut off
Prevents ageing of sensors when exposed to high temperatures
Communication
Options
USB host
Ethernet 10/100 Base-T
Size and weight
Wall and pipe mount
(H x D x W)
236.5 x 160 x 250 mm - weight 4.25 kg
9.31 x 6.30 x 9.84 in. - weight 8.82 lbs
Panel mount: Face (housing)
(H x D x W)
156 (123) x 250 x 220 (214) mm - weight 3.35 kg
6.14 (4.84) x 9.84 x 8.86 (8.43) in. - weight 6.62 lbs
Summary of Contents for Orbisphere 51 Series
Page 5: ...4 Table of Contents...
Page 19: ...18 Specifications...
Page 48: ...47 Section 6 View Menu Figure 33 View menu...
Page 52: ...51 Section 7 Measurement Menu Figure 36 Measurement menu...
Page 66: ...65 Section 9 Inputs Outputs Menu Figure 40 Inputs Outputs menu...
Page 68: ...67 Inputs Outputs Menu 9 4 Analog outputs Figure 41 Analog outputs menu...
Page 75: ...74 Inputs Outputs Menu...
Page 91: ...90 Communication Menu...
Page 94: ...93 Section 12 Products Menu Figure 44 Products menu...
Page 97: ...96 Global Configuration Menu...
Page 98: ...97 Section 14 Services menu Figure 46 Services menu Part 1...
Page 99: ...98 Services menu Figure 47 Services menu Part 2...
Page 103: ...102 Services menu...