20
Installation
3.2
Installation check list
To complete the installation, proceed to the following actions:
1.
Follow the installation instructions in this section carefully.
2.
On completion, verify once more that everything is properly connected.
3.
If using the instrument with a TC sensor it is very important to ensure the purge gas supply
has been connected properly and has been turned on, and that a purge safety backup unit
has been installed. Refer to
4.
Turn the instrument on and set the operating language.
5.
Enter the default login credentials “
1007
” for the ID and “
1234
” for the password. Refer to
.
6.
Change the default login and set the security levels, users ID’s and passwords. Refer to
7.
Configure the instrument and measurement channels.
8.
For the K1200 sensor, verify that the sensor cap selected (i.e. low or high range: K1200-L or
K1200-H) corresponds to the sensor in use. Refer to
.
9.
Enable thermal cutoff and set it to the minimum possible value (e.g. 10°C if the liquid in the
sample line is at 0-5°C). This is critical to ensure long-term stability. Refer to
.
10.
Set the measurement interval according to your response time needs. Refer to
.
11.
Enable the calibration timer and set the number of days to 180 days (6 months) or
according to your quality needs to ensure a regular sensor calibration. Refer to
.
12.
Enable the service timer and set the number of days to 360 days (1 year) or according to
your quality needs to ensure a regular sensor service. Refer to
13.
If required, program the relay and analog outputs to ensure an event is set when a
calibration or service is required. These can be programmed for each as custom events.
Refer to
. Note that when a calibration or service timer has
expired, the instrument will automatically display this information as a warning message.
Refer to
List of events and alarms on page 109
14.
Perform a barometric sensor calibration.
15.
Perform an external pressure calibration (optional).
16.
Set the alarm thresholds.
The instrument should now be ready for operation. If a problem should arise. Refer to
sensor maintenance on page 103
. If the difficulty cannot be overcome, please contact your
Hach representative who will be happy to assist you.
Summary of Contents for Orbisphere 51 Series
Page 5: ...4 Table of Contents...
Page 19: ...18 Specifications...
Page 48: ...47 Section 6 View Menu Figure 33 View menu...
Page 52: ...51 Section 7 Measurement Menu Figure 36 Measurement menu...
Page 66: ...65 Section 9 Inputs Outputs Menu Figure 40 Inputs Outputs menu...
Page 68: ...67 Inputs Outputs Menu 9 4 Analog outputs Figure 41 Analog outputs menu...
Page 75: ...74 Inputs Outputs Menu...
Page 91: ...90 Communication Menu...
Page 94: ...93 Section 12 Products Menu Figure 44 Products menu...
Page 97: ...96 Global Configuration Menu...
Page 98: ...97 Section 14 Services menu Figure 46 Services menu Part 1...
Page 99: ...98 Services menu Figure 47 Services menu Part 2...
Page 103: ...102 Services menu...