36
Installation
3.8.6
Weld-on stainless steel socket
The ORBISPHERE 29501 weld-on sensor socket can be used to install a sensor into a stainless
steel pipe (min.Ø 50 mm or 2”). When welding the socket to the pipe, check that setback
between the pipe’s inner diameter and the sensor tip does not exceed 4 mm (see diagram).
Note:
Be sure to remove the two O-rings from the socket before welding and leave the sensor’s stainless
steel cap screwed on during welding to prevent thread distortion.
Recommendation:
To facilitate sensor removal and installation, we suggest installing the socket in a location where
the liquid can be drained quickly and easily. By creating a one meter long piece of pipe
(
) with shut off valves at both ends, just a small volume of liquid needs to be drained to
enable sensor removal.
Also, a precise sensor and socket installation can be performed in the workshop, and this
assembly can be placed in the production line with minimal down time.
Figure 19 Weld-on sensor socket
1.
Pipe external diameter
2.
Setback allowed from internal pipe diameter
Figure 20 Installation in-line
1.
Shut off valve
4.
Shut off valve
2.
Sensor socket
5.
Sampling valve
3.
Sampling valve
Summary of Contents for Orbisphere 51 Series
Page 5: ...4 Table of Contents...
Page 19: ...18 Specifications...
Page 48: ...47 Section 6 View Menu Figure 33 View menu...
Page 52: ...51 Section 7 Measurement Menu Figure 36 Measurement menu...
Page 66: ...65 Section 9 Inputs Outputs Menu Figure 40 Inputs Outputs menu...
Page 68: ...67 Inputs Outputs Menu 9 4 Analog outputs Figure 41 Analog outputs menu...
Page 75: ...74 Inputs Outputs Menu...
Page 91: ...90 Communication Menu...
Page 94: ...93 Section 12 Products Menu Figure 44 Products menu...
Page 97: ...96 Global Configuration Menu...
Page 98: ...97 Section 14 Services menu Figure 46 Services menu Part 1...
Page 99: ...98 Services menu Figure 47 Services menu Part 2...
Page 103: ...102 Services menu...