40
Installation
The flow chamber should be mounted in such a way that the sample outlet port is located at the
lowest point to allow condensation to escape with the outgoing gas. Attach the flow chamber to
a vertical support with the screws supplied. The pressure sensor must be on top.
Note:
A user manufactured spacer (~15 mm thick) may be used between the flow chamber and support for
improved access for sensor removal.
Shown here with: gas sensor (right), pressure sensor (center) and optional second sensor (left)
Figure 24 32002.xxx multi parameter flow chamber
1.
Top view
5.
Front view
2.
Sensor
6.
Pressure sensor housing
3.
Optional second sensor
7.
Sample outlet
4.
Sample inlet