79
Communication Menu
User action log file example
The "User action log file" below contains 3 user actions.
Nr
mm/dd hh:mm:ss User ID
User Name
Action ID
Description
1 1/21
15:13:44
1007
Armstrong
140
Measurement config.
0 1/21
15:13:27
1007
Armstrong
132
Identification
2 1/21
15: 9:15
1007
Armstrong
132
Identification
Measurement file example
6 measurements are described below:
Nr mm/dd hh:mm:ss
Gas
Temp
Mask
Fluor. phi Barom Ext P. Index
[ppb]
[°C]
[°] [bar] [bar]
0
2/17
21:15:37
75.051
20.1
400
26.039 1.005 0.000 2271
1
2/17
21:15:27
75.043
20.1
400
26.045 1.005 0.000 2266
2
2/17
21:15:17
75.047
20.1
400
26.052 1.005 0.000 2261
3
2/17
21:14:57
75.044
20.1
400
26.041 1.005 0.000 2256
4
2/17
21:14:47
75.047
20.1
400
26.038 1.005 0.000 2251
5
2/17
21:14:37
75.050
20.1
400
26.054 1.005 0.000 2246
Configuration report example
INSTRUMENT CONFIGURATION
Measurement mode . . . . . . . . . Continuous mode
Pressure unit. . . . . . . . . . . . . . . [bar]
Temperature unit . . . . . . . . . . . . [°C]
Storage mode . . . . . . . . . . . . . . Rolling buffer
Storage RAM time. . . . . . . . . . . 10 [s]
Storage FLASH time . . . . . . . . . 3600 [s]
Channel 1
Membrane. . . . . . . . . . . . . . . . . 2956A
Medium . . . . . . . . . . . . . . . . . . . Liquid
Gas unit. . . . . . . . . . . . . . . . . . . mbar
Liquid . . . . . . . . . . . . . . . . . . . . Water
Resolution displayed. . . . . . . . . 3
Thermal cut off . . . . . . . . . . . . . Enabled60.0 [°C]
Alarms
Low Low . . . . . . . . . . . . . . . . . . Disabled0.000000 [mbar]
Low . . . . . . . . . . . . . . . . . . . . . . Disabled0.100000 [mbar]
High. . . . . . . . . . . . . . . . . . . . . . Enabled5.000000 [mbar]
High High . . . . . . . . . . . . . . . . . Disabled1.100000 [mbar]
Hysteresis . . . . . . . . . . . . . . . . . 1 [%]
Delay. . . . . . . . . . . . . . . . . . . . . 0 [s]
Filter State . . . . . . . . . . . . . . . . . Disabled
Type . . . . . . . . . . . . . . . . . . . . . Median
Depth . . . . . . . . . . . . . . . . . . . . 5
Central depth . . . . . . . . . . . . . . 3
Interferences
CO
2
or H
2
S . . . . . . . . . . . . . . . . All disabled
Chlorinity/Salinity . . . . . . . . . . . All disabled19.000000 [g/l]
H
2
. . . . . . . . . . . . . . . . . . . . . . . Disabled0.100000 [bar]
Summary of Contents for Orbisphere 51 Series
Page 5: ...4 Table of Contents...
Page 19: ...18 Specifications...
Page 48: ...47 Section 6 View Menu Figure 33 View menu...
Page 52: ...51 Section 7 Measurement Menu Figure 36 Measurement menu...
Page 66: ...65 Section 9 Inputs Outputs Menu Figure 40 Inputs Outputs menu...
Page 68: ...67 Inputs Outputs Menu 9 4 Analog outputs Figure 41 Analog outputs menu...
Page 75: ...74 Inputs Outputs Menu...
Page 91: ...90 Communication Menu...
Page 94: ...93 Section 12 Products Menu Figure 44 Products menu...
Page 97: ...96 Global Configuration Menu...
Page 98: ...97 Section 14 Services menu Figure 46 Services menu Part 1...
Page 99: ...98 Services menu Figure 47 Services menu Part 2...
Page 103: ...102 Services menu...