27
Installation
3.4.3
Connection panel (bottom of instrument)
Alternative instrument mounting procedure
When it is not convenient to work from the back of the panel, the instrument can be connected
before fitting in the panel.
1.
Install the panel support frame in the panel opening
2.
Slip the cables through the panel opening
3.
Remove the instrument cover
4.
Slip the cables through the instrument cover
5.
Slip the cables through the instrument back panel cable glands
6.
Connect the cables to the instrument electronic boards
7.
Tighten the cable glands
8.
Reinstall the instrument cover
9.
Install the instrument in the panel opening
1.
Power cable
2.
Ethernet cable gland
3.
USB-A host connector for mass storage
device
4.
USB-B 4 pin connector (reserved for
Hach service technicians)
5.
Input/Output 1 cable gland
6.
TC Sensor connection
7.
External pressure sensor connection
8.
Input/Output 2 cable gland
9.
K1200 Sensor connection
Figure 10 Panel mount connection panel
Summary of Contents for Orbisphere 51 Series
Page 5: ...4 Table of Contents...
Page 19: ...18 Specifications...
Page 48: ...47 Section 6 View Menu Figure 33 View menu...
Page 52: ...51 Section 7 Measurement Menu Figure 36 Measurement menu...
Page 66: ...65 Section 9 Inputs Outputs Menu Figure 40 Inputs Outputs menu...
Page 68: ...67 Inputs Outputs Menu 9 4 Analog outputs Figure 41 Analog outputs menu...
Page 75: ...74 Inputs Outputs Menu...
Page 91: ...90 Communication Menu...
Page 94: ...93 Section 12 Products Menu Figure 44 Products menu...
Page 97: ...96 Global Configuration Menu...
Page 98: ...97 Section 14 Services menu Figure 46 Services menu Part 1...
Page 99: ...98 Services menu Figure 47 Services menu Part 2...
Page 103: ...102 Services menu...