59
Calibration menu
8.1
K1200 sensor calibration
Sensor calibration
The sensor can be calibrated manually on an ad hoc basis. By default, the mode is set to zero calibration
with auto-end (refer to
Calibration configuration on page 60
for more details), but these parameters can
be changed.
For higher level concentrations (above 1% oxygen which corresponds to about 400 ppb dissolved O
2
) a
high level adjustment can be performed using a gas mixture containing more than 1% oxygen, or a
known line sample. However, this should not be done without first ensuring the zero point is accurate.
This can be achieved by performing a zero calibration first.
Low range sensors: (K1200-L spots)
There are two calibration modes available - zero or high level adjustment. The sensor is factory
calibrated at zero. During use, the zero calibration is the best calibration to guarantee the sensor
specifications. After a spot replacement, a zero calibration is recommended.
High range sensors: (K1200-H spots)
There are three calibration modes available - zero, high level adjustment or in 100% humid air.
The sensor is factory calibrated at zero and in 100% humid air. During use, humid air calibration is
the best calibration to guarantee the sensor specifications. After a spot replacement, a zero
calibration and a 100% humid air calibration are recommended.
Initial sensor calibration
The sensor has been calibrated at the factory prior to shipment and is ready for use upon delivery.
However, if the sensor has not been used for a period of more than six months since delivery, or if the
sensor spot has been replaced or changed in any way, then a sensor calibration will be required.
Select the
Configuration
option from the
Gas sensor
calibration menu to check the calibration
parameters.
Ensure the parameters are set up as illustrated left:
•
Auto-calibration
: This option is not available for the K1200 sensor.
•
Manual-calibration Auto-End
: On by default.
•
Hold during calibration or verification
: On by default.
•
Zero calibration bottle
: Ensure this is
disabled
by unchecking the box as it is not relevant for the
K1200 sensor.
Exit from the configuration screen by pressing on
OK
. Then select
Calibration
from the
Gas sensor
calibration menu and perform a manual zero calibration as described in
. For
high range sensors perform an additional 100% humid air calibration as described in
calibration (high range sensors only) on page 61
.
Manual calibration
Manual calibrations can be made at any time. Perform a calibration by following these steps:
1.
Remove the sensor from the sample line.
2.
Rinse the sensor head with clean water.
3.
Wipe the sensor head with a clean soft tissue to remove any excess moisture.
4.
If using the supplied calibration device (refer to
Portable calibration device - K1200 sensor on
), insert the sensor into the sensor holder on top of the calibration device. If not using the
calibration device, insert the sensor into the flow chamber.
5.
Flow the calibration sample through the calibration device or flow chamber as applicable. If using the
calibration device, fully open the valve on the pressure reducer to give a gas flow rate of 0.1 L/min. If
you are
not
using the supplied calibration device with pressure reducer, the
maximum
allowable
inlet pressure must be no more than 2 bar absolute.
6.
Configure the calibration as described in
Calibration configuration on page 60
.
7.
Start the calibration as described in
or
100% humid air calibration (high
range sensors only) on page 61
depending on the preferred calibration method.
Summary of Contents for Orbisphere 51 Series
Page 5: ...4 Table of Contents...
Page 19: ...18 Specifications...
Page 48: ...47 Section 6 View Menu Figure 33 View menu...
Page 52: ...51 Section 7 Measurement Menu Figure 36 Measurement menu...
Page 66: ...65 Section 9 Inputs Outputs Menu Figure 40 Inputs Outputs menu...
Page 68: ...67 Inputs Outputs Menu 9 4 Analog outputs Figure 41 Analog outputs menu...
Page 75: ...74 Inputs Outputs Menu...
Page 91: ...90 Communication Menu...
Page 94: ...93 Section 12 Products Menu Figure 44 Products menu...
Page 97: ...96 Global Configuration Menu...
Page 98: ...97 Section 14 Services menu Figure 46 Services menu Part 1...
Page 99: ...98 Services menu Figure 47 Services menu Part 2...
Page 103: ...102 Services menu...