77
Communication Menu
10.1.1 Data available
All individual data are separated by at least one tabulation character (ASCII code=0x09).
For the cyclic measurements, the data format is detailed. For the files, only one example for
each file is given to explain the data format.
Cyclic measurements
1.
If the option "Mode = standard" is chosen, the following message is sent for each active
channel:
with:
\t ................................................the ASCII tab character: code=0x09
\r ................................................the ASCII Carriage Return character: code=0x0D
\n ..............................................the ASCII Line Feed character: code=0x0A
CHn ..........................................the 2 ASCII characters "CH" + the channel no (from 1 to 3).
Gas ............................................the gas concentration.
Gas Unit ....................................the gas unit.
Temperature ..............................the temperature.
Temperature Unit.......................the temperature unit.
Barometric Pressure..................the barometric pressure.
Barometric Pressure Unit ..........the barometric pressure unit.
Event .........................................the event bit mask in hexadecimal format.
The values are not described here. Refer to
List of events and alarms on page 109
.
•
Example of one measurement:
CH1 697.176 mbar 20.1 °C 0.982 bar C00
2.
If the option "Mode = expert" is chosen, the following message is sent:
with:
Current ......................................The current measured in [microA].
Partial pressure .........................The partial pressure in [bar].
External Pressure .....................The external pressure.
External Pressure Unit...............The external pressure unit.
Time ..........................................The time of the measurement. Format "hh:mm:ss."
Index..........................................This is the index of the last measurement.
This number starts at 0 at power up of the program.
•
Example of one measurement:
CH1 697.173 mbar 20.1 °C 0.982 bar C00 80.056229 µA 0.697
bar 1.000 bar 12:59:42 5923
CHn\t Gas\t Gas Unit\t Temperature\t Temperature Unit\t Barometric Pressure\t Barometric Pressure Unit\t\ Event\t\r\n
CHn\t Gas\t Gas Unit\t Temperature\t Temperature Unit\t BarometricPressure\t Barometric Pressure Unit\t
Event\t Current\t µA\t Partial pressure\t bar\t External Pressure\t External Pressure Unit\t Time\t Index\r\n
Summary of Contents for Orbisphere 51 Series
Page 5: ...4 Table of Contents...
Page 19: ...18 Specifications...
Page 48: ...47 Section 6 View Menu Figure 33 View menu...
Page 52: ...51 Section 7 Measurement Menu Figure 36 Measurement menu...
Page 66: ...65 Section 9 Inputs Outputs Menu Figure 40 Inputs Outputs menu...
Page 68: ...67 Inputs Outputs Menu 9 4 Analog outputs Figure 41 Analog outputs menu...
Page 75: ...74 Inputs Outputs Menu...
Page 91: ...90 Communication Menu...
Page 94: ...93 Section 12 Products Menu Figure 44 Products menu...
Page 97: ...96 Global Configuration Menu...
Page 98: ...97 Section 14 Services menu Figure 46 Services menu Part 1...
Page 99: ...98 Services menu Figure 47 Services menu Part 2...
Page 103: ...102 Services menu...