67
Inputs/Outputs Menu
Tri-linear
Range
Measurement M
Resolution R
4-20 mA
1: AOL > I > 4
M=MLL+(ML-MLL)
*
(I-4)/(AOL-4)
R=(ML-MLL)
*
20/((AOL-4)
*
1010)
2: AOH > I > AOL
M=ML+(MH-ML)
*
(I-AOL)/ (AOH-AOL)
R=(MH-ML)
*
20/((AOH-AOL)
*
1010)
3: 20 > I > AOH
M=MH+(MHH-MH)
*
(I-AOH) / (20-AOH) R=(MHH-MH)
*
20/((20-AOH)
*
1010)
0-20 mA
1: AOL > I > 0
M=MLL+(ML-MLL)
*
I/AOL
R=(ML-MLL)
*
20/(AOL
*
1010)
2: AOH > I > AOL
M=ML+(MH-ML)
*
(I-AOL)/ (AOH-AOL)
R=(MH-ML)
*
20/((AOH-AOL)
*
1010)
3: 20 > I > AOH
M=MH+(MHH-MH)
*
(I-AOH)/(20-AOH)
R=(MHH-MH)
*
20/((20-AOH)
*
1010)
Analog output characteristics (continued)
None
This is the default value.
Setting the analog output to “None” means that the output value will always
be zero and importantly ensures that no current is emitted on that output, so
reducing power consumption as well as reducing heat within the instrument.
Summary of Contents for ORBISPHERE K1200
Page 5: ...4 Table of Contents...
Page 13: ...12 General Information...
Page 39: ...38 Installation...
Page 45: ...44 User Interface...
Page 46: ...45 Section 5 View Menu Figure 34 View menu...
Page 49: ...48 View Menu...
Page 59: ...58 Calibration Menu...
Page 60: ...59 Section 8 Inputs Outputs Menu Figure 40 Inputs Outputs menu...
Page 62: ...61 Inputs Outputs Menu 8 4 Analog outputs Figure 41 Analog outputs menu...
Page 69: ...68 Inputs Outputs Menu...
Page 86: ...85 Section 11 Products Menu Figure 44 Products menu...
Page 89: ...88 Global Configuration Menu...
Page 90: ...89 Section 13 Services Menu Figure 46 Services menu Part 1...
Page 91: ...90 Services Menu Figure 47 Services menu Part 2...