pH, Low Ionic Strength (LIS) Water
(50 to 150
µ
S/cm)
For drinking water and natural waters
1.
Install the
Potassium Chloride
Electrolyte Gel
Cartridge in the
Platinum Series pH
Electrode. (See
Section
for
assembly instructions.)
Note: The Hach
sens
ion
™2
meter is
used for this analysis.
2.
Prime the electrode
by pushing the
dispenser button until
gel comes out of the
reference junction.
Rinse excess gel from
the tip and the outlet.
3.
Connect the
electrode to the
sens
ion
2
pH/ISE
meter.
Note: Ensure that the
electrode has been
conditioned according
to instructions in
Section 1.3.2 on
page 12.
Note: Ensure meter is
set up for 5-pin
connector.