Table of contents
on page 3
on page 7
Additional information
Additional information is available on the manufacturer's website.
Specifications
Specifications are subject to change without notice.
Specification
Details
Dimensions bench space (W x D x H) 2.72 x 0.83 x 0.58 m (9 x 2.75 x 2 ft) (recommended)
Location
Laboratory (indoor use only)
Pollution degree/installation category
II
Protection class
I
Operating temperature
5 to 40 °C (41 to 104 °F), maximum 80% relative humidity non-
condensing for temperatures up to 31 ºC and decreases linearly to 50%
relative humidity at 40 ºC
Storage temperature
–20 to 70 °C (–4 to 158 °F)
Altitude limit
Up to 3000 m (10,000 feet) above sea level
Certifications
CE: IEC/EN 61010-1 per LV Directive (2006/95/EC), IEC/EN 61326 per
EMC: Directive (2004/108/EC)
cETL: CSA C22.2 No 61010-1
ETLus: UL 61010-1
Warranty
1 year
System unit (5 channels)
Dimensions (W x D x H)
70.5 x 71.2 x 26.7 cm (27.7 x 28.0 x 10.5 in.)
Weight
46.5 kg (102.51 lb)
Power requirements
100–115 V ±10%, 230 V ±10%; 50–60 Hz, 800 W
Fuse rating
T, 8 A, 100–115 V, T, 4 A, 230 V
Lamp
Tungsten/Halogen, 12 V, 50 W, MR16
Sample processing module (SPM)
Dimensions (W x D x H)
69 x 9 x 26 cm (27.2 x 3.5 x 10.2 in.)
Weight
4 kg (8.8 lb)
Power requirements
115 VAC, 1.2 A, 60 Hz or 230 VAC, 0.8 A, 50 Hz
Fuse rating
T, 1.6 A, 250 V, 5 mm x 20 mm
Pump (12 position)
Dimensions (W x D x H)
15 x 26 x 26.7 cm (5.9 x 10.2 x 10.5 in.)
Weight
5.6 kg (12.3 lb)
English