1.
From the main menu use the or and and to select
MEASURE. Confirm.
2.
Use the to change the following settings. Confirm every entry.
Option
Description
Resolution
Select the resolution: 1, 0.1, 0.01 (default) or 0.001
Measure
Stability—select By stability Criterion: Fast (variation <
0.02 pH in 6 s), Standard (variation < 0.01 pH in 6 s) or Strict
(variation < 0.002 pH in 6 s). In continuous—enter the time
interval for the In continuous Acquis. interval (data storage or
printing data). By time—enter the time interval for data storage
or printing data.
Display mV
Display mV—select YES or NO to show mV.
Limits
Limits—select YES or NO. YES: Enter the upper and lower
limits. An acoustic warning appears when the measurement is
out of limit. The report output shows an A next to the
measured value when the measurement was out of limit.
Isopotential
Isopotential—change the Isopotential pH value in Data
introduction. Select Calculate to calibrate the probe again.
3.
Push to start the measurement.
Note: If the measurement is not stabilizing after 120 seconds, the meter turns
automatically into the continuous measurement mode.
Section 7 Advanced operation
7.1 Change the measurement units
The measurement units can be changed individually for each channel.
1.
From the main menu use the or to select SYSTEM. Confirm.
2.
Use the or to select Measurement units and confirm.
3.
Select Channel 1 or Channel 2 and confirm.
4.
Select ORP (mV), pH or ISE and confirm.
7.2 Use a sample ID
The sample ID tag is used to associate readings with a particular sample
location. If assigned, stored data will include this ID.
1.
From the main menu use the or to select SYSTEM. Confirm.
2.
Use the or to select Sample ID and confirm.
3.
Use the or to select
Option
Description
Automatic
A consecutive number will be automatically assigned to every
sample.
Manual
A keyboard or a barcode scanner is required to enter the
sample ID name before taking a measurement (maximum
15 characters).
7.3 Select the data output
Data can be stored or transferred to a printer or to a PC.
1.
From the main menu use the or to select SYSTEM. Confirm.
2.
Use the or to select Data Output and confirm.
3.
Use the or to select
Option
Description
Deactivated
Select Deactivated if no printer or PC is connected.
For Printer
Select Dot matrix printer or Thermal printer.
For Computer
Select Terminal, LabCom or LabCom Easy. The LabCom
Software controls several modules, pH and conductivity
meters, automatic burettes, Samplers and so on from a
computer. The LabCom Easy software gets measurements
and calibration data from the meters
7.4 Change the date and time
The date and time can be changed from the Date / Time menu.
English
11
Summary of Contents for sensION MM340
Page 2: ...Table of Contents English 3 Fran ais 20 Espa ol 38 Portugu s 56 74 90 108 125 2...
Page 6: ...1 2 3 4 6 English...
Page 15: ...1 2 English 15...
Page 23: ...1 2 3 4 Fran ais 23...
Page 33: ...1 2 Fran ais 33...
Page 41: ...1 2 3 4 Espa ol 41...
Page 51: ...1 2 Espa ol 51...
Page 59: ...1 2 3 4 Portugu s 59...
Page 69: ...1 2 Portugu s 69...
Page 76: ...1 1 5 2 pH 4 01 pH 7 00 pH 10 01 6 o 3 7 4 8 3 3 1 76...
Page 77: ...1 2 3 4 77...
Page 78: ...3 2 1 2 2 3 3 4 2 1 1 1 7 PC DIN 2 2 8 1 3 1 9 pH 1 4 1 10 pH 2 5 PC RS 232 11 2 2 6 3 78...
Page 85: ...9 9 1 9 2 87 9 3 85...
Page 86: ...1 2 86...
Page 91: ...2 1 1 2 1 2 2 2 ICES 003 A A FCC PART 15 A FCC 15 1 2 FCC 15 A 1 2 3 4 5 91...
Page 93: ...1 2 3 4 93...
Page 95: ...4 4 1 1 5 2 6 3 7 1 2 4 1 2 4 2 ID ID 4 1 ID 4 2 pH ORP mV ISE 5 F 3 6 4 3 5 5 1 AC 5 2 95...
Page 102: ...9 9 1 9 2 104 9 3 102...
Page 103: ...1 2 103...
Page 110: ...sensION MM340 2 pH ORP mV ISE PC 2 4 1 1 1 5 2 pH 4 01 pH 7 00 pH 10 01 6 3 7 4 8 3 3 1 110...
Page 111: ...1 2 3 4 111...
Page 120: ...1 2 120...
Page 121: ...10 1 119 100 pH EC DO 240 ISE 119 3 0 C 1 119 DO DO 119 EC 121...
Page 128: ...1 2 3 4 128...
Page 136: ...9 1 9 2 138 9 3 136...
Page 137: ...1 2 137...
Page 138: ...10 1 136 100 s pH EC DO 240 s ISE 136 3 0 C 1 136 DO DO EC 136 EC 138...
Page 142: ......
Page 143: ......