Table of Contents
4 User interface and navigation
11 Replacement parts and accessories
Section 1 Specifications
Specifications are subject to change without notice.
Specification
Details
Dimensions
35 x 20 x 11 cm (13.78 x 7.87 x 4.33 in.)
Weight
1100 g (2.43 lb)
Meter enclosure
IP42, indoor use
Power requirements
(external)
100–240 V, 0.4 A, 47-63 Hz
Meter protection class
Class III
Installation category
Category I
Altitude requirements
Standard 2000 m (6562 ft) ASL (Above Sea Level)
Storage temperature
–15 to +65 °C (5 to +149 °F)
Operating temperature
0 to 40 °C (41 to 104 °F)
Operating humidity
< 80% (non-condensing)
Resolution
pH: 0.1/0.01/0.001, ORP: 0.1/1 mV, ISE:
programmable, temperature: 0.1 ºC (0.18 ºF)
Measuring error (± 1 digit)
pH: ≤ 0.005, ORP: ≤ 0.2 mV, temperature: ≤
0.2 ºC (≤ 0.36 ºF)
Specification
Details
Reproducibility (± 1 digit)
pH: ± 0.001, ORP: ± 0.1 mV, temperature:
± 0.1 ºC (± 0.18 ºF)
Data storage
330 results and last 9 calibrations
Connections
2 Combined or indicator probe: BNC connector
(Imp. >10
12
Ω); 2 Reference electrodes: banana
connector; A.T.C. type Pt 1000: banana or
telephonic; 2 magnetic stirrers: RCA connector
Temperature correction
Manual, Pt 1000 temperature probe (A.T.C.),
isopotential pH programmable, standard value
7.00
Measurement display lock
Continuous measurement, by stability
Display
Liquid crystal, backlit, 128 x 64 dots
Keyboard
PET with protective treatment
Certification
CE
Section 2 General information
In no event will the manufacturer be liable for direct, indirect, special,
incidental or consequential damages resulting from any defect or
omission in this manual. The manufacturer reserves the right to make
changes in this manual and the products it describes at any time, without
notice or obligation. Revised editions are found on the manufacturer’s
website.
2.1 Safety information
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or
misuse of this product including, without limitation, direct, incidental and
consequential damages, and disclaims such damages to the full extent permitted
under applicable law. The user is solely responsible to identify critical application
risks and install appropriate mechanisms to protect processes during a possible
equipment malfunction.
English
3
Summary of Contents for sensION MM340
Page 2: ...Table of Contents English 3 Fran ais 20 Espa ol 38 Portugu s 56 74 90 108 125 2...
Page 6: ...1 2 3 4 6 English...
Page 15: ...1 2 English 15...
Page 23: ...1 2 3 4 Fran ais 23...
Page 33: ...1 2 Fran ais 33...
Page 41: ...1 2 3 4 Espa ol 41...
Page 51: ...1 2 Espa ol 51...
Page 59: ...1 2 3 4 Portugu s 59...
Page 69: ...1 2 Portugu s 69...
Page 76: ...1 1 5 2 pH 4 01 pH 7 00 pH 10 01 6 o 3 7 4 8 3 3 1 76...
Page 77: ...1 2 3 4 77...
Page 78: ...3 2 1 2 2 3 3 4 2 1 1 1 7 PC DIN 2 2 8 1 3 1 9 pH 1 4 1 10 pH 2 5 PC RS 232 11 2 2 6 3 78...
Page 85: ...9 9 1 9 2 87 9 3 85...
Page 86: ...1 2 86...
Page 91: ...2 1 1 2 1 2 2 2 ICES 003 A A FCC PART 15 A FCC 15 1 2 FCC 15 A 1 2 3 4 5 91...
Page 93: ...1 2 3 4 93...
Page 95: ...4 4 1 1 5 2 6 3 7 1 2 4 1 2 4 2 ID ID 4 1 ID 4 2 pH ORP mV ISE 5 F 3 6 4 3 5 5 1 AC 5 2 95...
Page 102: ...9 9 1 9 2 104 9 3 102...
Page 103: ...1 2 103...
Page 110: ...sensION MM340 2 pH ORP mV ISE PC 2 4 1 1 1 5 2 pH 4 01 pH 7 00 pH 10 01 6 3 7 4 8 3 3 1 110...
Page 111: ...1 2 3 4 111...
Page 120: ...1 2 120...
Page 121: ...10 1 119 100 pH EC DO 240 ISE 119 3 0 C 1 119 DO DO 119 EC 121...
Page 128: ...1 2 3 4 128...
Page 136: ...9 1 9 2 138 9 3 136...
Page 137: ...1 2 137...
Page 138: ...10 1 136 100 s pH EC DO 240 s ISE 136 3 0 C 1 136 DO DO EC 136 EC 138...
Page 142: ......
Page 143: ......