stabilization time. To change the measurement parameter (if applicable),
push and hold pH or ORP (mV).
To measure the sample continuously, push pH or ORP (mV) to start a
measurement and push pH or ORP (mV) again during the stabilization.
The parameter flashes to indicate the continuous measurement mode.

Advanced operation

Change the date and time

The date and time can be changed from the Date and Time menu.

1.

Use the 

 to enter the Date and Time menu. The date and time

format: dd-mm-yyyy 24h.
The attribute to be changed will be automatically highlighted and will
advance to the next attribute if no 

 or 

 is pushed.

2.

To change an attribute, push 

 or 

 when the highlighted

attribute is selected.
The current date and time will be shown on the display.

Adjust the display contrast

1.

Push the 

 and 

 at the same time to enter the Display contrast

menu.

2.

Use the 

 or 

 to adjust the contrast of the display. The meter

returns automatically to the measurement screen after 3 seconds.

Change the stirring settings

The stirring speed can be changed during calibration and during a
measurement.

1.

Push 

 during a calibration or during a measurement to enter the

stirring menu.

2.

Use the 

 or 

 to change the stirring speed in %.

Change the temperature units

The temperature units can be changed to Celsius or Fahrenheit.

1.

From the main screen push 

.

2.

Push 

 to select between Celsius or Fahrenheit.

Maintenance

C A U T I O N

 

Multiple hazards. Only qualified personnel must conduct the tasks
described in this section of the document.

N O T I C E

 

Do not disassemble the instrument for maintenance. If the internal components
must be cleaned or repaired, contact the manufacturer.

Clean the instrument

N O T I C E

 

Never use cleaning agents such as turpentine, acetone or similar products to
clean the instrument including the display and accessories.

Clean the exterior of the instrument with a moist cloth and a mild soap
solution.

Clean the probe

Clean the probe as needed. Refer to 

Troubleshooting

 on page 12 for

more information about cleaning. Refer to the probe documentation for
information about the probe maintenance.
Use the cleaning agents listed in 

Table 1

 for contaminations on the pH

probe.

English

   

9

Summary of Contents for sensION+ pH3

Page 1: ...utilisation Manual del usuario Manual do utilizador U ivatelsk p ru ka Brugsanvisning Gebruikershandleiding Instrukcja obs ugi Bruksanvisning K ytt j n k sikirja Haszn lati tmutat Manual de utilizare...

Page 2: ...27 Fran ais 39 Espa ol 51 Portugu s 63 e tina 75 Dansk 87 Nederlands 99 Polski 111 Svenska 123 Suomi 135 147 Magyar 160 Rom n 172 lietuvi kalba 184 196 T rk e 208 Slovensk jazyk 219 Slovenski 231 Hrva...

Page 3: ...General information Revised editions are found on the manufacturer s website Safety information N O T I C E The manufacturer is not responsible for any damages due to misapplication or misuse of this...

Page 4: ...opean domestic or public disposal systems Return old or end of life equipment to the manufacturer for disposal at no charge to the user Product overview The sensION meters are used with probes to meas...

Page 5: ...1 2 3 4 English 5...

Page 6: ...1 Select the correct adapter plug for the power outlet from the adapter kit 2 Connect the universal power adapter to the meter Figure 2 3 Connect the universal power adapter to an AC receptacle Figur...

Page 7: ...ample temperature C or F 2 Measurement mode or time and date 4 Visual measurement timer Navigation Use the calibrate key to calibrate the probe Use the parameter key to take a sample measurement Use t...

Page 8: ...ration solution is shown 7 Rinse the probe with deionized water and put the probe into the third calibration beaker Be sure that there are no air bubbles in the membrane 8 Push to measure the third ca...

Page 9: ...ngs The stirring speed can be changed during calibration and during a measurement 1 Push during a calibration or during a measurement to enter the stirring menu 2 Use the or to change the stirring spe...

Page 10: ...eaning agent Proteins Pepsin cleaning solution Grease oils fats Electrode cleaning solution Limescale 0 1 N HCl solution Replace the magnetic stirrer If the magnetic stirrer does not start follow the...

Page 11: ...1 2 English 11...

Page 12: ...conditions Examine the probe Clean the probe refer to Clean the probe on page 9 for more information make sure that there are no air bubbles in the membrane Shake the probe like a thermometer connect...

Page 13: ...mL LZW9463 99 pH buffer solution 7 00 250 mL LZW9464 98 pH buffer solution 10 01 250 mL LZW9471 99 pH buffer solution 4 01 1000 mL LZW9466 99 pH buffer solution 7 00 1000 mL LZW9467 98 Consumables co...

Page 14: ...8 10 17 245 20 68 2 00 4 00 7 02 9 26 10 06 228 25 77 2 00 4 01 7 00 9 21 10 01 220 30 86 2 00 4 01 6 99 9 16 9 96 212 40 104 2 00 4 03 6 97 9 06 9 88 195 50 122 2 00 4 06 6 97 8 99 9 82 178 60 140 2...

Reviews: