Resolution pH:
0.1
0.01
0.001
Measurement condit.
Meas.:
By stability
In Continuous
By time
Measurement condit.
Meas.:
By stability
In Continuous
By time
Measurement condit.
Meas.:
By stability
In Continuous
By time
Measurement condit.
Resolution: 0.01
Meas.: By stability
Display mV: No
Limits: No
Isopotential: pH 7
Measurement condit.
Isopotencial:
pH 7
Data introduction
Calculate
Measurement condit.
Resolution: 0.01
Meas.: By stability
Display mV: No
Limits: No
Isopotential: pH 7
Measurement condit.
Display mV:
Yes
No
Measurement condit.
Resolution: 0.01
Meas.: By stability
Display mV: No
Limits: No
Isopotential: pH 7
Measurement condit.
Resolution: 0.01
Meas.: By stability
Display mV: No
Limits: No
Isopotential: pH 7
Measurement condit.
Measurement condit.
By Stability
Criteria
Rapid
Standard
Strict
Measurement condit.
By Stability
Criteria
Rapid
Standard
Strict
Measurement condit.
In Continuous
Acquis. Interval
01
:00
Measurement condit.
By time
Acquis. Interval
01:
30
Measurement condit.
In Continuous
Acquis. Interval
01:
30
Measurement condit.
By time
Measurement time
00:
01
:00
Resolution: 0.01
Meas.: By stability
Display mV: No
Limits: No
Isopotential: pH 7
Measurement condit.
Limits:
Yes
No
Measurement condit.
Measurement condit.
Limits
Upper: pH 8.20
Lower: pH 6.05
Exit
Measurement condit.
Lower= pH 6.05
Upper limit?
pH
8.
20
...
Start Measurement
...by stability
10-04-2010 17:15
pH MEASURE
New calibration
in 1 day
Default programming (standard)
Other options
Resolution
•
0.01 pH
• 0.1 pH
• 0.01 pH
• 0.001 pH
Measuring mode
•
By stability
.
- Standard criterion (variation < 0 .01 pH in 6 s) .
•
In continuous
.
- The result is stored when the measurement
has finished .
- The result is printed when the measurement
has finished .
•
By stability.
- Rapid criterion (variation < 0 .02 pH in 6 s) .
- Standard criterion (variation < 0 .01 pH in 6 s) .
- Strict criterion (variation < 0 .002 pH in 6 s) .
•
In continuous
. Selection of the time interval for:
- Data storage .
- Data printing .
•
By time
. Selection of the time interval for:
- Data storage .
- Data printing .
View mV
•
No
• Simultaneous display of pH and mV concerned.
Limits
•
No
• Higher and lower limits. Acoustic warning when the measure-
ment is out of limits, in the printed ticket appears “A” beside
the measured value .
pH isopotencial
•
pHiso 7.00
pHiso 7 .00 / Manual introduction / Value calculation .
“Re-programming” flow chart
4. Operation
Important note:
Once the measurement has been re-programmed, it will be performed in
accordance with the new parameters .
It is highly recommended to apply the same stability criteria to measure and
calibrate .
4.8. Re-programming pH measurement
11
ENGLISH
Data storage in
the Data Logger. It
coincides with the
printing and PC
sending. Selecting
0 store/print at
the end of the
measurement.
Start measurement
See Electrode isopotential on page 18.
Start of
measurement with
new programming.
etc.