8.
Deliver the first volume by pressing the trigger. The display shows the next volume.
9.
Continue dispensing by repeating step 3. After the last step,
Blowout
is displayed.
10.
To empty the tip completely, press the trigger.
11.
Change the tip and continue pipetting.
Note: The pipette can be emptied at any time by pressing Cancel (left selection key, refer to
on page 7).
Mix + Dilute
1.
Select
Program
>
Edit
, then select with the scroll key
Mix + Dilute
and accept with
OK
.
2.
Select the pipetting first volume by pressing the scroll key up or down.
3.
Accept the volume with
OK
.
4.
Select the second volume (
VOL 2
) with the scroll key and accept with
OK
.
5.
Select the pipetting speeds.
6.
Fill a clean reagent reservoir with the liquid to be dispensed.
7.
Dip the tip under the surface of the first liquid in the reservoir and press the trigger. The first
volume is drawn in to the tip.
Air
is displayed.
8.
Withdraw the tip from the liquid touching it against the edge of the reservoir to remove excess
liquid. Press the trigger again to aspirate the air buffer.
9.
Dip the tip under the surface of the second liquid and press the trigger. The second volume is
drawn in to the tip.
10.
Withdraw the tip from the liquid.
11.
To deliver the both volumes, press the trigger.
Mix
is displayed.
12.
By pressing the trigger the pipette starts to pipette approximately 70% of the total volume several
times as long as the trigger is held pressed.
13.
After releasing the trigger, the pipette stops after the next dispensing and
Blowout
is displayed.
14.
Withdraw the tip from the liquid and press the trigger and hold it down to empty the tip.
15.
Release the trigger to return to the ready position.
16.
Change the tip and continue pipetting.
Manual
With manual mode it is possible to measure volumes. In manual mode only slower speeds are
available to ensure a quick stop.
1.
Select
Program
>
Edit
, then select with the scroll key
Manual
and accept with
OK
.
2.
Then select a limit for the total volume, the default is the maximum volume.
3.
Select the pipetting speeds.
0.00
is then displayed.
4.
Fill a clean reagent reservoir with the liquid to be dispensed.
5.
Dip the tip under the surface of the liquid in the reservoir and press the trigger. The liquid is drawn
in to the tip and the actual volume is shown on the display.
6.
The pipetting direction can be changed with the
In/Out
key (left selection key, refer to
on page 7).
7.
To deliver the liquid, select the
Out
key (down direction) and press the trigger.
8.
To deliver the rest of the liquid, press the trigger and hold it down and hold it down to empty the
tip completely.
9.
Change the tip and continue pipetting.
Note: The volume display can be reset to zero at any time by pressing Reset with the right selection key. Refer
to
on page 7.
English
11
Summary of Contents for TenSette plus
Page 2: ...English 3 Deutsch 17 Fran ais 33 Nederlands 49 2...
Page 65: ......