10.
If the default stirring speed needs to be adjusted, push the up and down arrow keys to increase or
decrease the speed.
Note: This adjustment only applies to the current operation. The standard default stirring speed for the
application is not changed.
11.
Two options are available during the procedure:
Option Description
Stop
Aborts the calibration and no results are calculated. If selected during the
Replicate Sample
option,
then all data in the series is lost.
Skip
Stops the current operation and goes directly to the next step in the procedure. Results are then
calculated from calibration data available before
Skip
is selected. Results can be less accurate
using this option.
12.
Push
Reject
or
Continue
.
Option
Description
Reject
Rejects the calibration. Select
Cancel
to go back to the result display or
Confirm
to reject the
calibration. If this is the first calibration, select
Confirm
to reject the calibration and use the default
or previous calibration values. If this is a
Replicate Standard
calibration, select
Confirm
to reject
only the current calibration in the series.
Continue
Select from one of these options:
•
Replicate Standard
: Do the calibration again using the same standard
•
Save & Exit
: Keep the calibration results and exit the calibration procedure
•
Reject & Exit
: Reject the calibration results and use the default or previous calibration values,
and exit the calibration procedure
Auto leveling calibration
This option is only available when at least one installed application contains an auto leveling
calibration method. The calibration makes sure that the sample volume in the measurement cell is
the same before each titration. An external pump must be installed for this procedure to extract
sample from the beaker.
Refer to the Full User Manual, which is available as a download from our website.
Get a sample measurement
Use this option to get sample measurements with one of the installed applications.
1.
From the home screen, select the measurement application, then push
Select
. Application
information shows on the display.
2.
Read the related "Application Note" from the USB applications key for more instructions.
3.
If necessary, select an icon for more information or to change some data.
Option
Description
Information
Shows more information about the application.
Operator
Changes the operator ID. Select from a list of applicable operators.
Sample
Sample Name: Change the specified name of the sample.
Type: Push the left and right arrow keys and select the sample type (
Sample
,
QC
or
Define
blank
) to be used for the measurement. If
Define blank
has been selected before, two more
sample types are available (
QC with blank
and
Sample with blank
).
4.
Fill a beaker with the recommended sample amount that shows on the display. If necessary, add
more of the solvent specified in the application note until the sensor is correctly installed in the
sample.
5.
Carefully put a magnetic stir bar into the beaker. Make sure there is no liquid spill.
6.
Attach the beaker to the sensor holder.
English
21
Summary of Contents for TitraLab AT1000 Series
Page 10: ...10 English ...
Page 32: ...32 Deutsch ...
Page 77: ...Français 77 ...
Page 144: ...144 Čeština ...
Page 166: ...166 Dansk ...
Page 188: ...188 Nederlands ...
Page 232: ...232 Svenska ...
Page 254: ...254 Suomi ...
Page 299: ...Magyar 299 ...
Page 321: ...Română 321 ...
Page 343: ...Türkçe 343 ...
Page 365: ...Slovenský jazyk 365 ...
Page 387: ...Slovenski 387 ...
Page 409: ...Hrvatski 409 ...
Page 431: ...Ελληνικά 431 ...
Page 447: ......