1
Temperature sensor
2
Reference sensor
3
Measuring sensor
3.
The legacy adapter commissioning wizard automatically starts. Follow the instructions that show
on the display. Select the parameter based on the connected sensor.
Option
Description
pH
Select this parameter if the connected sensor is a pH analogic sensor.
Metal/RedOX/Color
Select this parameter if the connected sensor is a Pt-Pt (metallic) analogic sensor or
PTM450/OPT300 sensor.
ISE
Select this parameter if the connected sensor is an ion selective sensor.
Refer to the application documentation to enter the related information for the selected
parameter.
Refer to the full user manual for additional information about how to reset the Legacy adapter or
change the sensor name.
Connect the sensor
Use a conical adapter to hold the sensor tightly in the sensor holder.
Connect the sensor to an available sensor port on the rear of the instrument. After the sensor is
connected, make sure that the sensor icon shows in the banner at the top of the display.
N O T I C E
Make sure that the sensor tip is 5 to 10 mm above the top of the magnetic stir bar to prevent any contact with the
bar during operation.
Refer to the illustrated steps that follow.
English
13
Summary of Contents for TitraLab AT1000 Series
Page 10: ...10 English ...
Page 32: ...32 Deutsch ...
Page 77: ...Français 77 ...
Page 144: ...144 Čeština ...
Page 166: ...166 Dansk ...
Page 188: ...188 Nederlands ...
Page 232: ...232 Svenska ...
Page 254: ...254 Suomi ...
Page 299: ...Magyar 299 ...
Page 321: ...Română 321 ...
Page 343: ...Türkçe 343 ...
Page 365: ...Slovenský jazyk 365 ...
Page 387: ...Slovenski 387 ...
Page 409: ...Hrvatski 409 ...
Page 431: ...Ελληνικά 431 ...
Page 447: ......