4.
Push
Import
to install the selected applications.
5.
Push
OK
to complete the installation. The installed applications show on the home screen.
Note: To install more applications, push Home to go to the home screen, then remove the USB key and
reconnect it.
Prepare the instrument for measurement
1.
From the home screen, select
Purge
. All attached devices are listed.
2.
Select
All elements
to purge all the attached devices, or select one device to purge. Push
Select
. Air is removed from the device and filled with liquid from the bottle.
3.
Push
OK
when the operation has completed.
4.
Make sure that there are no air bubbles in the device. Do step
again if there are any air
bubbles.
5.
Select the next device to purge if individual devices are being selected.
6.
Push
Exit
when all the tubes are filled with reagent and the device has no air bubbles.
Note: If a few small air bubbles can be seen on the inner wall and/or piston of the syringe, they can be left
without effecting system performance.
Standard operations
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
C A U T I O N
Personal injury hazard. Never use the instrument without the syringe protection cover in place.
C A U T I O N
Chemical exposure hazard. Never remove the stir bar from the beaker before the end of a titration.
Calibration
Calibrate the sensor
1.
From the home screen, select
Calibration
, then push
Electrode calibration
.
2.
If more than one sensor is installed, push the up and down arrow keys to highlight the sensor to
use, then push
Select
.
3.
If more than one application includes calibration parameters for the sensor, push the up and
down arrow keys to highlight the application to use, then push
Select
. Calibration information
shows on the display.
4.
If necessary, select an icon for more information or to change some data.
Option
Description
Electrode
Shows more information about the sensor.
English
19
Summary of Contents for TitraLab AT1000 Series
Page 10: ...10 English ...
Page 32: ...32 Deutsch ...
Page 77: ...Français 77 ...
Page 144: ...144 Čeština ...
Page 166: ...166 Dansk ...
Page 188: ...188 Nederlands ...
Page 232: ...232 Svenska ...
Page 254: ...254 Suomi ...
Page 299: ...Magyar 299 ...
Page 321: ...Română 321 ...
Page 343: ...Türkçe 343 ...
Page 365: ...Slovenský jazyk 365 ...
Page 387: ...Slovenski 387 ...
Page 409: ...Hrvatski 409 ...
Page 431: ...Ελληνικά 431 ...
Page 447: ......