7.
Make sure that the icon at the bottom of the display
is highlighted. Do the instructions that
show on the display adjacent to this icon. Refer to
on page 13 to make sure
that the tubes and sensor are correctly aligned.
8.
Push
Start
to start the measurement. Measurement data shows on the display.
9.
If the default stirring speed needs to be adjusted, push the up and down arrow keys to increase or
decrease the speed.
Note: This adjustment only applies to the current operation. The standard default stirring speed for the
application is not changed.
10.
Two options are available during the procedure:
Option Description
Stop
Aborts the measurement and no results are calculated. If selected during the
Replicate Sample
option, then all data in the series is lost.
Skip
Stops the current operation and goes directly to the next step in the procedure. Results are then
calculated from measurement data available before
Skip
is selected. Results can be less accurate
using this option.
11.
When the measurement is complete, push the arrow keys to see the different measurement
views.
12.
Push
Next
for these options:
Option
Description
Replicate
Sample
Use this option to start the same titration on the same sample. This is used to study the
repeatability by successively analyzing several parts of the same sample. At the end of
each measurement, a window shows the average value, the standard deviation and the
relative standard deviation.
New Sample
Use this option to start the same titration on a new sample. No standard deviation or
relative standard deviation measurements will be done.
13.
Push
Exit
to go back to the home screen.
Syringe verification
Use the syringe verification option to examine the syringe accuracy based on the ISO 8655 standard.
Note: A license key is necessary to enable the syringe verification option on the instrument. Contact the
manufacturer or a sales representative to get a license key.
1.
From the home screen, select Settings, then push
Select
.
2.
From the settings screen, select Options, then push
Select
.
3.
Push the left or right arrow key to select the expert mode, then push
Done
.
4.
From the home screen, select Maintenance, then push
Select
.
5.
From the maintenance screen, select Syringe verification, then push
Select
.
6.
When the option is selected for the first time, enter the license key.
7.
Follow the procedure supplied with the syringe verification kit. For additional information, refer to
the expanded version of this manual.
Manage the data log
To select data to view, delete or export, specify data filters
1.
From the home screen, select
Data log
.
2.
Select an option, then push
Select
.
Option
Description
View data log
Views measurement data. Select individual lines of data to view more content.
22
English
Summary of Contents for TitraLab AT1000 Series
Page 10: ...10 English ...
Page 32: ...32 Deutsch ...
Page 77: ...Français 77 ...
Page 144: ...144 Čeština ...
Page 166: ...166 Dansk ...
Page 188: ...188 Nederlands ...
Page 232: ...232 Svenska ...
Page 254: ...254 Suomi ...
Page 299: ...Magyar 299 ...
Page 321: ...Română 321 ...
Page 343: ...Türkçe 343 ...
Page 365: ...Slovenský jazyk 365 ...
Page 387: ...Slovenski 387 ...
Page 409: ...Hrvatski 409 ...
Page 431: ...Ελληνικά 431 ...
Page 447: ......