16
Installation
Figure 6
Electrical connection of the magnetic stirrer
3.4.3
Installation of tubes
Make sure that the tubes are never kinked.
Regularly check the system as well as all tubes and connections for leaks, condition and
cleanliness.
To prevent twisting of the tubes, you should always connect flange tube connections
before those with grommets.
The system is delivered with two tubes; these are supplied pre-installed on the TitraLab
and only need to be connected to the relevant titrant bottle and to the electrode.
3.4.3.1
Removal of protective stickers (Uninstalled tubes configuration)
1.
Remove the protective stickers on the burette valve.
2.
Make sure that the pre-installed white O-rings are not removed.
3.
Check that the O-rings are in a horizontal position inside the thread.
1
Connect the cable of the magnetic stirrer to the
STIRRER
port.
WA R N I N G
Risk of injury to eyes, skin and respiratory
During the first days of system use, check the screw joint between the syringe and valve to avoid
leaks and the potential escape of reagents.
When working with chemicals and/or solvents, always observe the relevant accident prevention
regulations and wear appropriate personal protective clothing for eye, face, hand, body and/or
respiratory protection.