26
Technical Data
0572 Argos²
total weight: 5,6kg
dimension: W314xH135xD223 mm
power supply: 230 V ~, 50 Hz
P
max
: 300 W
fuse primary:
F1, F2: 500 mA delay action fuse
F3, F4: 3,15 A delay action fuse
fuse Sec. Thermo fuse
protection class I
accuracy of speed display: ±10 %
handpiece:
type B applied part
diameter: 17-22 mm
length: 152 mm
speed: 7000-40000 rpm
Accessories & Spare Parts
art. no.
0970
foot switch
3497
spray liquid filter
4571
washer
5116
filter
5179
micro filter
6158
cleaning brush
8555
AquaHy
The clamping system is designed
for rotating instruments, which have
a shaft of 2,35 mm in accordance
with DIN EN ISO 1797-1.
Operating Time
The device is designed for extend-
ed usage
.
Ambient Conditions
The device is intended for use in
dry closed rooms.
Operation:
Temp.: +10°C to +26°C
Rel. humidity: 25% to 75%
(not condensed)
Air pressure: 700 hPa to 1060 hPa
Storage:
Temp: -5°C to +55°C
Rel. humidity: 10% to 95%
(not condensed)
Air pressure: 500 hPa to 1060 hPa
Transport (up to 4 weeks):
Temp: -5°C to +55°C
Rel. humidity: 10% to 95%
(not condensed)
Air pressure: 500 hPa to 1060 hPa
Declaration of Conformity
We declare under our own respon-
sibility, that the following foot care
unit 0572 are consistent with the
following guidelines:
2014/30/EU EMC guide line
2014/35/EU low voltage guide line
hadewe GmbH
–
Grambartstraße 10
30165 Hannover
–
Germany
Safety Check
A safety check must be done in
compliance with the regulations of
the country where the device is
used. The inspection interval is
once a year.