30
Technical Data
Guidance and Manufacturer's Declaration
–
Electromagnetic Immunity IEC 60601-1-2
The device is suitable for use in the specified electromagnetic environment. The customer
and/or the user of the device should assure that it is used in an electromagnetic environment
as described below:
Immunity Test
IEC 60601 Test Level
Compliance
Level
Electromagnetic Environ-
ment
–
Guidance
Conducted RF IEC
61000-4-6
Radiated RF IEC
61000-4-3
3 Vrms 150 kHz to 80
MHz
3 V/m
80 MHz to 2,5 GHz
3 Veff
3 V/m
Portable and mobile RF
communications equipment
should be used no closer to
any part of the pa-on, includ-
ing cables, than the recom-
mended separation distance
calculated from the equation
appropriate for the frequen-
cy of the transmitter.
Recommended Separation
Distance:
d = 1,17 √P
d= 1,17 √P 80 MHz to 800
MHz
d = 2,3 √P 800 MHz to 2,5
GHz
where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from fixed
RF transmitters, as deter-
mined by an electromagnetic
site survey, a should be less
than the compliance level in
each frequency range
Interference may occur in
the vicinity of equipment:
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is af-
fected by absorption and reflection from structures, objects, and people.